FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,11.5

MDR report key: 16766292 · Received April 18, 2023

Report

Report Number
0002023141-2023-01109
Event Type
Injury
Date Received
April 18, 2023
Date of Event
February 24, 2023
Report Date
October 4, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019232
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR IMPLANT INFECTION HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORD REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH DID NOT SUGGEST THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/ OR ANY FURTHER ESCALATIONS. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1260102. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD OP# (B)(4) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1260102 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THE INFECTION OF AN IMPLANT HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT TOOTH #7 EXTRACTED WITH IMMEDIATE IMPLANT PLACEMENT. PATIENT RETURNED FOR POST OP EXAM WITH COMPLAINTS OF PAIN, SWELLING AND BAD TASTE. THE IMPLANT WAS INFECTED AND REMOVED. SURGICAL/MEDICAL INTERVENTION REQUIRED FOR PERMANENT DAMAGE: YES, IMPLANT REMOVED & DEBRIDED. ADDITIONAL APPOINTMENT REQUIRED: NO. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS, PAIN AND INFLAMMATION.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1207271 IMP,TSV,4.1MM,SBM,11.5 DENTAL IMPLANT DZE ZIMMER DENTAL TSV4B11 1260102 00889024019232

Patients

Seq Age Sex Outcome Treatment
1 76 YR Prefer Not To Disclose Required Intervention