FDA Adverse Event Malfunction Summary report: N

TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE

MDR report key: 16765986 · Received April 18, 2023

Report

Report Number
9610773-2023-01073
Event Type
Malfunction
Date Received
April 18, 2023
Report Date
July 19, 2023
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
EWY
UDI-DI
04042761052764
PMA / PMN Number
K912362
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED AS THE BLACK COLORED EYEPIECE FUNNEL WAS FOUND BROKEN/FRACTURED AND SEPARATED FROM THE SCOPE. THE INSPECTION ALSO NOTED THE RIGID OUTER TUBE HAS DEFORMATION/DENTED (NON-REPORTABLE). THE REVIEW OF THE DEVICE HISTORY RECORDS FOR THE AFFECTED LOT OR SERIAL NUMBER WITHOUT SHOWING ANY NON-CONFORMITIES OR DEVIATIONS REGARDING THE DESCRIBED ISSUE. THE DEVICE WAS MANUFACTURED ACCORDING TO VALID INSTRUCTIONS AND MET ALL SPECIFICATIONS. THE INVESTIGATION IS STILL IN PROGRESS; HOWEVER, IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY. IN GENERAL, THE END-USER IS REQUIRED TO INSPECT THE DEVICE FOR DEFECTS, CHECK THE FUNCTION OF ALL DEVICES AND HAVE ALTERNATE EQUIPMENT PRIOR TO USE. ADDITIONAL 510(K): K923982/ K950076.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 8 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY BROKEN EYEPIECE OCCURRED DUE TO THE USE OF EXCESSIVE FORCE BY THE CUSTOMER. HOWEVER, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE HIGH-DEFINITION QUICK LOCK AUTOCLAVABLE TELESCOPE WAS RETURNED TO THE OLYMPUS REGIONAL REPAIR CENTER FOR A REPORTED ¿EYEPIECE OUT OF ORDER¿. THE CUSTOMER REPORTED PROBLEM WAS FOUND AT AN UNSPECIFIED EVENT. DURING INSPECTION, THE BLACK EYEPIECE FUNNEL WAS FOUND BROKEN/FRACTURED FROM THE SCOPE. NO DEATH OR INJURY AND NO IMPACT TO PATIENT OR OTHER HAS BEEN REPORTED TO OLYMPUS. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE BROKEN EYEPIECE FUNNEL FOUND DURING REPAIR INSPECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1188187 TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE RIGID SCOPE EWY OLYMPUS WINTER & IBE GMBH WA53005A 667476 04042761052764

Patients

Seq Age Sex Outcome Treatment
1 Unknown