KIT, RAD-97 PULSE CO-OXIMETER
Report
- Report Number
- 3019388613-2023-00128
- Event Type
- Malfunction
- Date Received
- April 18, 2023
- Date of Event
- March 22, 2023
- Report Date
- March 22, 2023
- Manufacturer
- MASIMO - 15750 ALTON PKWY
- Product Code
- MWI
- UDI-DI
- 00843997012317
- PMA / PMN Number
- K193626
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ADDITIONAL MANUFACTURING NARRATIVE: ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED., OTHER, OTHER TEXT: INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: (B)(6). INITIAL REPORTER PHONE NUMBER EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, PHONE NUMBER IS AS FOLLOWS: (B)(6). DEVICE NOT RETURNED.
ADDITIONAL MANUFACURING NARRATIVE: OTHER TEXT: THE RETURNED DEVICE WAS INVESTIGATED. THE DEVICE BATTERY WAS NOT RECORDING THE FULL CHARGE CAPACITY AS IT DEPLETED QUICKLY AND SHUT DOWN WITH A LOW BATTERY ERROR MESSAGE. THE DEVICE WAS THEN UNABLE TO BE POWERED BACK ON USING THE BATTERY. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, ZIP CODE IS AS FOLLOWS: G51 4TF. INITIAL REPORTER PHONE NUMBER EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS, PHONE NUMBER IS AS FOLLOWS: 0141 452 3286.
THE CUSTOMER REPORTED THE DEVICE POWERED OFF DURING USE AND WON'T POWER BACK ON. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
THE CUSTOMER REPORTED THE DEVICE POWERED OFF DURING USE AND WON'T POWER BACK ON. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1186879 | KIT, RAD-97 PULSE CO-OXIMETER | MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) | MWI | MASIMO - 15750 ALTON PKWY | 27572 | 00843997012317 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |