TOMOTHERAPY TREATMENT DELIVERY SYSTEM
MDR report key: 16765568
·
Received April 18, 2023
Report
- Report Number
- 3003873069-2023-00007
- Event Type
- Malfunction
- Date Received
- April 18, 2023
- Date of Event
- March 23, 2023
- Report Date
- January 9, 2024
- Manufacturer
- ACCURAY INCORPORATED
- Product Code
- IYE
- PMA / PMN Number
- K121934
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
CUSTOMER ALLEGED THERE WAS A COUCH FAULT. WHEN A PATIENT LAID DOWN ON THE COUCH, THE Z+ BUTTON WAS USED TO RAISE THE COUCH, HOWEVER, WHEN THE Z+ BUTTON WAS RELEASED, THE COUCH CONTINUED TO RAISE TO THE HIGHEST POSITION BEFORE STOPPING. THE CCK (COUCH CONTROL KEYPAD) BUTTONS WOULD NOT STOP THE COUCH MOVEMENT. NO INJURY WAS REPORTED. INVESTIGATION IS ONGOING.
Additional Manufacturer Narrative · 0
THE INVESTIGATION CONCLUDED THAT THE CAUSE OF THE EVENT WAS DUE TO THE HARDWARE. IN ORDER TO RESOLVE THE ISSUE, THE CCA (COUCH CONTROL ASSEMBLY) WAS REPLACED. NO HARM WAS CAUSED TO THE PATIENT. THE RISK WAS DETERMINED TO BE ACCEPTABLE.
Description of Event or Problem · 0
THERE WAS A REPORT OF A COUCH FAULT, THE CCK (COUCH CONTROL KEYPAD) COULD NOT STOP THE COUCH MOVEMENT.