FDA Adverse Event Malfunction Summary report: N

AED3

MDR report key: 16763122 · Received April 18, 2023

Report

Report Number
1220908-2023-01322
Event Type
Malfunction
Date Received
April 18, 2023
Report Date
March 28, 2023
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION FOR EVALUATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A FAULTY MAIN BOARD. THE MAIN BOARD WAS REPLACED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE EMITTED AN ARC UPON POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1210792 AED3 DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED3 NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown