HAHN TAPERED IMPLANT Ø5.0 X 10 MM
Report
- Report Number
- 3011649314-2023-00223
- Event Type
- Injury
- Date Received
- April 18, 2023
- Date of Event
- March 20, 2023
- Report Date
- April 7, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE HAS BEEN RETURNED. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.
CORRECTION: SECTION A1: PATIENT INITIALS CORRECTED FROM (B)(6). THE DEVICE HAS BEEN RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT# 6120947 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR LOT # 6120947 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø5.0 X 10 MM (70-1154-IMP0015) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. BONE DEBRIS WAS OBSERVED IN THE THREADING OF THE IMPLANT, (SEE ATTACHED IMAGES). THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE: "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 4989 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU 4989 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 4989 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN. THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.
CORRECTION SECTION B1 AND B2 REPORTABILITY UPDATED TO A SERIOUS INJURY BASED ON THE INFORMATION PROVIDED. SECTION H1: UPDATED TO A SERIOUS INJURY BASED ON THE INFORMATION PROVIDED. THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.
CAPA CA-00016. MANUFACTURER REFERENCE: (B)(4).
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS II, AND THEIR ORAL HYGIENE IS LISTED AS GOOD. THERE IS NO MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT PLACEMENT. THE PATIENT PRESENTED ON (B)(6) 2022 FOR A PRIMARY PROCEDURE ON TOOTH #19. THE PATIENT RETURNED ON 03-20-23, AFTER THE FINAL PROSTHESIS DELIVERY, WITH COMPLAINTS OF DISCOMFORT AND SENSITIVITY. UPON EXAMINATION, THE PROVIDER NOTED THAT THE IMPLANT FAILED TO INTEGRATE, AND THE DEVICE WAS REMOVED. BASED ON THE DATES NOTED IN THE QUESTIONNAIRE COMPLETED BY THE PROVIDER, A MEDICAL OPINION WAS SOUGHT, AND IT WAS DETERMINED THAT BECAUSE THE IMPLANT WAS PLACED AND REMOVED IN SUCH A SHORT TIME, IT LACKED STABILITY.
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS II, AND THEIR ORAL HYGIENE IS LISTED AS GOOD. THERE IS NO MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT PLACEMENT. THE PATIENT PRESENTED ON (B)(6) 2022 FOR A PRIMARY PROCEDURE ON TOOTH #19. THE PATIENT RETURNED ON (B)(6) 2023, AFTER THE FINAL IMPRESSION FOR PROSTHESIS, WITH COMPLAINTS OF DISCOMFORT AND SENSITIVITY. UPON EXAMINATION, THE PROVIDER NOTED THAT THE IMPLANT FAILED TO INTEGRATE, AND THE DEVICE WAS REMOVED AND NOT REPLACED. THE SYMPTOMS RESOLVED AFTER IMPLANT REMOVAL, THERE WAS NO PERMANENT PATIENT INJURY AND NO ADDITIONAL MEDICAL/SURGICAL PROCEDURE WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1092076 | HAHN TAPERED IMPLANT Ø5.0 X 10 MM | HAHN TAPERED IMPLANT | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0015 | 6120947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male | Required Intervention |