FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 16758945
·
Received April 18, 2023
Report
- Report Number
- 9610877-2023-53614
- Event Type
- Malfunction
- Date Received
- April 18, 2023
- Date of Event
- March 20, 2023
- Report Date
- April 18, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- GCL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT. THEREFORE, 510K IS NOT APPLICABLE. MODEL EE-1580K IS AVAILABLE IN THE USA, WITH A 510K NUMBER K961564. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED, THAT THE CCD MODULE WITH DRIVE PCB BLACKOUT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED, DUE TO THE PHYSICAL DAMAGE APPLIED ON THE CCD MODULE WITH DRIVE PCB. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 624110 | PENTAX | VIDEO ESOPHAGOSCOPE | GCL | HOYA CORPORATION PENTAX TOKYO OFFICE | EE17-J10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |