FDA Adverse Event Malfunction Summary report: N

BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM

MDR report key: 16756443 · Received April 17, 2023

Report

Report Number
8041187-2023-00161
Event Type
Malfunction
Date Received
April 17, 2023
Date of Event
March 12, 2023
Report Date
May 26, 2023
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 01-MAY-2023 H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2231205. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY A SAMPLE WAS RETURNED TO OUR FACILITY TO AID IN OUR INVESTIGATION. OUR ENGINEERS NOTED THAT THE QSYTE COMPONENT WAS CONTAMINATED WITH A SMALL BLACK FOREIGN MATERIAL. CLOSER INSPECTION OF THE MATERIAL FURTHER IDENTIFIED THE MATERIAL AS A FIBER OR HAIR LIKE STRUCTURE THAT MOST LIKELY ORIGINATED FORM THE PACKAGING PERSONNEL.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: WHEN OPENED FOR USE, A FOREIGN OBJECT WAS FOUND IN THE Q-SYTE CONNECTOR.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: WHEN OPENED FOR USE, A FOREIGN OBJECT WAS FOUND IN THE Q-SYTE CONNECTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2320107 BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 2231205

Patients

Seq Age Sex Outcome Treatment
1 Unknown