FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 16754527 · Received April 17, 2023

Report

Report Number
1000113657-2023-00216
Event Type
Malfunction
Date Received
April 17, 2023
Date of Event
March 23, 2023
Report Date
April 17, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007829
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 161, 244 AND 226 MG/DL. ERRATIC (BACK TO BACK) BLOOD GLUCOSE TEST RESULTS WERE NOT PROVIDED. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 150-160 MG/DL AND EXPECTED PM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 90-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 04/19/2023 AND TEST STRIPS WERE OPENED A FEW WEEKS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE NOT SET): RESULT 1: 185 MG/DL , DATE:(B)(6), TIME: 11:46 AM FASTING. RESULT 2: 112MG/DL , DATE: (B)(6),TIME: 4:19 PM FASTING. RESULT 3: 161MG/DL , DATE: (B)(6), TIME: 9:56 AM FASTING. RESULT 4: 244 MG/DL, DATE: (B)(6), TIME: 10:08 AM FASTING. RESULT 5: 226 MG/DL , DATE: (B)(6), TIME: 9:07 PM FASTING. RESULT 6: 88 MG/DL , DATE: (B)(6), TIME: 4:43 PM UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1808162 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX WGN 25CT24/CS MG/DL #383237 ZY4505S 00021292007829

Patients

Seq Age Sex Outcome Treatment
1 Unknown