XD DUAL FLAT SCREEN (24) HOLDER
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 9710055-2023-00309
- Event Type
- Malfunction
- Date Received
- April 17, 2023
- Date of Event
- April 12, 2023
- Report Date
- April 17, 2023
- Manufacturer
- MAQUET SAS
- Product Code
- FXR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ACCORDING TO THE REPORTING TIMEFRAME WE WOULD LIKE TO PROVIDE THE INFORMATION ABOUT CURRENT STATUS OF THE ISSUE. PLEASE BE ADVISED THAT IT IS BEING INVESTIGATED. ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. THE CORRECTION OF B5 DESCRIBE EVENT AND PROBLEM DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS B5 DESCRIBE EVENT AND PROBLEM: ON 12TH APRIL, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE MONITOR HANDLE HOLDER CAME OFF FROM THE MONITOR. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE SERIOUS INJURY. CORRECTED B5 DESCRIBE EVENT AND PROBLEM: ON 12TH APRIL, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EQUIPMENT - XD DUAL FLAT SCREEN (24) HOLDER. IT WAS STATED THE MONITOR HANDLE HOLDER CAME OFF FROM THE MONITOR. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE SERIOUS INJURY. THE CORRECTION OF D1 BRAND NAME AND D2A PRODUCT CODE DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS D1 BRAND NAME: POWERLED. CORRECTED D1 BRAND NAME: XD DUAL FLAT SCREEN (24) HOLDER. PREVIOUS D2A PRODUCT CODE: FSY. CORRECTED D2A PRODUCT CODE: FXR.
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EQUIPMENT - XD DUAL FLAT SCREEN (24) HOLDER. IT WAS STATED THE MONITOR HANDLE HOLDER CAME OFF FROM THE MONITOR. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE SERIOUS INJURY. BASED ON THE INFORMATION COLLECTED, IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, MAQUET EQUIPMENT DID NOT MEET ITS SPECIFICATION, DUE HANDLE HOLDER WHICH CAME OFF FROM THE MONITOR. SUCH A SCENARIO COULD BE CONSIDERED AS TECHNICAL DEFICIENCY, AND IN THIS WAY THE DEVICE CONTRIBUTED TO THE EVENT. PROVIDED INFORMATION DOES NOT INDICATE IF UPON THE EVENT OCCURRENCE, THE DEVICE WAS OR WAS NOT BEING USED FOR PATIENT TREATMENT. WHEN REVIEWING SIMILAR REPORTABLE EVENTS FOR THE SAME DEVICE TYPE, IT WAS CONFIRMED THAT IN THE LAST 5 YEARS, REGISTERED FOR THE ISSUE OF HANDLE HOLDER DETACHMENT ON MAQUET EQUIPMENT THERE WAS NO SERIOUS INJURY OR WORSE REPORTED. AS STATED BY THE SUBJECT MATTER EXPERT AT MANUFACTURING SITE, THE DETACHMENT BETWEEN HANDLE HOLDER AND ALUMINUM CYLINDER, WHICH ALLOWS BONDING OF THE HANDLE, IS PROBABLY DUE TO MECHANICAL SHOCKS, COLLISIONS OR EXCESSIVE EFFORTS. TO PREVENT ANY SIMILAR INCIDENT, THE FOLLOWING HAVE BEEN IMPLEMENTED: DURING THE MANUFACTURING THE PARTS ARE DEGREASED; THE QUANTITY OF THE INSTANT ADHESIVE GEL DEPOSITED IS CHECKED; THE ADHESIVE BONDING IS CHECKED BY MANUAL TRACTION DURING THE MANUFACTURING; THE USER MANUAL (IFU 01821 EN 11) ON PAGES 35 AND 40 MENTIONS TO CHECK THE CONDITION OF THE STERIZABLE HANDLE. WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER¿S RECOMMENDATION HAD BEEN FOLLOWED THE INCIDENT COULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. EVENT SITE NAME: (B)(6). H3 OTHER TEXT: DEVICE NOT RETURNED TO MANUFACTURER.
ON 12TH APRIL, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EQUIPMENT - XD DUAL FLAT SCREEN (24) HOLDER. IT WAS STATED THE MONITOR HANDLE HOLDER CAME OFF FROM THE MONITOR. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE SERIOUS INJURY.
MANUFACTURER'S REFERENCE NUMBER (B)(4).
ON 12TH APRIL, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE MONITOR HANDLE HOLDER CAME OFF FROM THE MONITOR. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS OR PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1053870 | XD DUAL FLAT SCREEN (24) HOLDER | HOLDER, CAMERA, SURGICAL | FXR | MAQUET SAS | ARD567701940 | ||
| 1827340 | XD DUAL FLAT SCREEN (24) HOLDER | HOLDER, CAMERA, SURGICAL | FXR | MAQUET SAS | ARD567701940 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |