FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1675074
·
Received May 5, 2010
Report
- Report Number
- 9611820-2010-00007
- Event Type
- Injury
- Date Received
- May 5, 2010
- Date of Event
- March 23, 2010
- Report Date
- April 6, 2010
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
ON APRIL 1, 2010, THE DOCTOR REPORTED THAT AN ABUTMENT FRACTURED. AN EVALUATION OF THE RETURNED PRODUCT INDICATED THAT THE ABUTMENT HAD BEEN CUSTOMIZED FOR THE PATIENT BY REDUCING ITS SIZE, MAKING IT UNABLE TO SUPPORT THE SUPERSTRUCTURE. AS A RESULT, THE SUPERSTRUCTURE BECAME LOOSE AND APPLIED ADDITIONAL STRESS ON THE ABUTMENT, CAUSING IT TO FRACTURE. THIS IS NOT A PRODUCT FAILURE.
Description of Event or Problem · 1
ON APRIL 6, 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS (B) (4) THAT A PITT EASY ABUTMENT FRACTURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH | 844201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Other| R |