FDA Adverse Event Malfunction Summary report: N

BD KIESTRA READA COMPACT

MDR report key: 16748648 · Received April 16, 2023

Report

Report Number
3010141591-2023-00002
Event Type
Malfunction
Date Received
April 16, 2023
Date of Event
March 16, 2023
Report Date
July 5, 2023
Manufacturer
BD KIESTRA LAB AUTOMATION
Product Code
JTQ
UDI-DI
00382904472061
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6. INVESTIGATION SUMMARY: THE CUSTOMER REPORTED ON THE READA COMPACT MATERIAL NUMBER: (B)(4) SERIAL NUMBER: (B)(6). IT WAS REPORTED BY THE CUSTOMER THAT 3 DISHES THAT HAS MOLD ON IT AND RC3 HAS A LOT OF HUMIDITY AND BAD SMELL. BD TECHNICAL SERVICES RECOMMENDED THAT THE CUSTOMER PERFORM THE REQUIRED MAINTENANCE TO CLEAN THE INCUBATOR. A FIELD SERVICE ENGINEER (FSE) WENT ONSITE AND REPLACED THE HEPA FILTER. AFTER THIS, THE ISSUE HAS BEEN RESOLVED AND NOW THE INSTRUMENT IS BACK TO ROUTINE USE. THE COMPLAINT IS CONFIRMED. NO NEW TRENDS, RISKS, OR HAZARDS WERE IDENTIFIED AS A RESULT OF THIS COMPLAINT. NO PATIENT IMPACT WAS REPORTED AS PART OF THIS COMPLAINT. DEVICE HISTORY RECORD (DHR) REVIEW IS NOT REQUIRED FOR THIS COMPLAINT. THE COMPLAINT WAS EVALUATED VIA OTHER ELEMENTS OF THE INVESTIGATION. THE RESULTS OF THIS EVALUATION HAVE NOT IDENTIFIED ANY NEW HAZARDS, NEW RISKS, OR SPECIFIC TRENDS. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS ASSOCIATED WITH THIS ISSUE. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD KIESTRA READA COMPACT THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ¿ DURING THE READING CUSTOMER FOUND THAT THERE IS 3 DISHES THAT HAS MOLD ON IT . ¿ CUSTOMER GAVE THE DISHES NUMBER TO PROCEED TO FIND ALL THE LOCATIONS THAT WERE ON THE RC. ¿ CUSTOMER PROCEED TO DECONTAMINATE ALL THE AREAS WHEN THIS DISHES WERE. ¿ CUSTOMER IS ALSO COMPLAINING THAT THIS RC3 HAS A LOT HUMIDITY AND BAD SMELL. ¿ CUSTOMER IS ASKING FROM COMPLETE CHECK OF THIS RC AND ASKING DECONTAMINATION PROCEDURE. ¿ FOR NOW CUSTOMER WILL CHECK PLATE BY PLATE TO ENSURE THAT THERE IS NO MOLD ON IT. R 2. ISSUE DESCRIPTION: CUSTOMER REPORTS THAT FIND MOLD IN THE RC3 IN 3 DISHES IN RC3.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD KIESTRA READA COMPACT THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING THE READING CUSTOMER FOUND THAT THERE IS 3 DISHES THAT HAS MOLD ON IT. CUSTOMER GAVE THE DISHES NUMBER TO PROCEED TO FIND ALL THE LOCATIONS THAT WERE ON THE RC. CUSTOMER PROCEED TO DECONTAMINATE ALL THE AREAS WHEN THIS DISHES WERE. CUSTOMER IS ALSO COMPLAINING THAT THIS RC3 HAS A LOT HUMIDITY AND BAD SMELL. CUSTOMER IS ASKING FROM COMPLETE CHECK OF THIS RC AND ASKING DECONTAMINATION PROCEDURE. FOR NOW CUSTOMER WILL CHECK PLATE BY PLATE TO ENSURE THAT THERE IS NO MOLD ON IT. ISSUE DESCRIPTION: CUSTOMER REPORTS THAT FIND MOLD IN THE RC3 IN 3 DISHES IN RC3.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2259828 BD KIESTRA READA COMPACT NOT CLASSIFIED JTQ BD KIESTRA LAB AUTOMATION 447206 00382904472061

Patients

Seq Age Sex Outcome Treatment
1 Unknown