FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE

MDR report key: 16747007 · Received April 14, 2023

Report

Report Number
1920898-2023-00227
Event Type
Malfunction
Date Received
April 14, 2023
Date of Event
March 27, 2023
Report Date
July 27, 2023
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K190054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 08-MAY-2023. H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (1) 1/2 ML SYRINGE IN AN OPEN POLYBAG FROM LOT# 2038577. IT WAS REPORTED THAT SYRINGE IN A BAG THAT WAS CRUSHED/DAMAGED NO NEEDLE. RETURNED SAMPLE VISUALLY EXAMINED AND OBSERVED DAMAGED SYRINGE WITH MISSING CANNULA HUB. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 2020067. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER INDICATED FAILURE. ROOT -CAUSE: NA H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED WHEN HE REMOVED THE NEEDLE SHIELD, THE NEEDLE AND PART THAT NEEDLE IS ATTACHED TO CAME OFF.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED WHEN HE REMOVED THE NEEDLE SHIELD, THE NEEDLE AND PART THAT NEEDLE IS ATTACHED TO CAME OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2280002 BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMF BD MEDICAL - DIABETES CARE 2020067

Patients

Seq Age Sex Outcome Treatment
1 Unknown