BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Report
- Report Number
- 1920898-2023-00227
- Event Type
- Malfunction
- Date Received
- April 14, 2023
- Date of Event
- March 27, 2023
- Report Date
- July 27, 2023
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- K190054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 08-MAY-2023. H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (1) 1/2 ML SYRINGE IN AN OPEN POLYBAG FROM LOT# 2038577. IT WAS REPORTED THAT SYRINGE IN A BAG THAT WAS CRUSHED/DAMAGED NO NEEDLE. RETURNED SAMPLE VISUALLY EXAMINED AND OBSERVED DAMAGED SYRINGE WITH MISSING CANNULA HUB. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 2020067. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER INDICATED FAILURE. ROOT -CAUSE: NA H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED WHEN HE REMOVED THE NEEDLE SHIELD, THE NEEDLE AND PART THAT NEEDLE IS ATTACHED TO CAME OFF.
IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED WHEN HE REMOVED THE NEEDLE SHIELD, THE NEEDLE AND PART THAT NEEDLE IS ATTACHED TO CAME OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2280002 | BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMF | BD MEDICAL - DIABETES CARE | 2020067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |