AWL, MICROFRACTURE, 45 DEGREES
Report
- Report Number
- 1017294-2010-00057
- Event Type
- Malfunction
- Date Received
- April 28, 2010
- Date of Event
- March 29, 2010
- Report Date
- March 29, 2010
- Manufacturer
- CONMED LINVATEC
- Product Code
- HWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION RESULTS: CONMED LINVATEC RECEIVED THIS AWL FOR EVAL. A VISUAL EXAMINATION CONFIRMED THE REPORTED PROBLEM AND FOUND THE AREA OF BREAKAGE BENT IN THE WRONG DIRECTION. THIS TYPE OF DAMAGE IS SEEN IF THE DEVICE IS USED AS A LEVER OR EXCESSIVE FORCE IS APPLIED DURING USE. A 2 YEAR REVIEW OF DEVICE HISTORY SHOWS NO OTHER REPORTS FOR THIS FAILURE MODE. THE INFO FOR USE (IFU) PROVIDES THE FOLLOWING PRECAUTIONS: INSPECT INSTRUMENT PRIOR TO USE TO ENSURE IT IS IN GOOD PHYSICAL CONDITION AND FUNCTIONS PROPERLY. THERE SHOULD BE NO LOOSE, BROKEN OR MISALIGNED PARTS. EXERCISE CARE IN THE USE OF THE DEVICE TO MINIMIZE SIDE OR BENDING LOADS. DO NOT USE EXCESSIVE FORCE ON INSTRUMENTS TO AVOID DAMAGE OR BREAKAGE DURING USE. AVOID UNINTENDED CONTACT WITH OTHER SURGICAL INSTRUMENTS DURING USE TO PREVENT DAMAGE OR BREAKAGE. IN ADDITION, THE USER IS INFORMED TO ALWAYS INSPECT AND TEST THE INSTRUMENT FOR PROPER FUNCTION PRIOR TO USE.
THE CUSTOMER REPORTED THAT DURING USE OF THIS AWL TO PERFORM AN ARTHROSCOPIC HIP PROCEDURE, THE TIP OF THIS INSTRUMENT BROKE INSIDE THE SURGICAL SITE. THE TIP WAS RETRIEVED BY USING A GRASPER FORCEPS, WHICH RESULTED IN A DELAY OF ABOUT 10 MINUTES. THE PROCEDURE WAS COMPLETED AS INTENDED WITHOUT INJURY TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AWL, MICROFRACTURE, 45 DEGREES | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | HWJ | CONMED LINVATEC | 00118908 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |