FDA Adverse Event Malfunction Summary report: N

TEMPERATURE/FLOW PROBE

MDR report key: 1674508 · Received April 30, 2010

Report

Report Number
9611451-2010-00258
Event Type
Malfunction
Date Received
April 30, 2010
Report Date
April 1, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE #2: LOT# 090618; DATE OF MFR: 06/18/2009. BACKGROUND: THE TEMPERATURE/FLOW PROBE IS A REUSABLE DEVICE INSERTED INTO THE BREATHING CIRCUIT WHICH SENDS TEMPERATURE AND FLOW INFO BY WAY OF ELECTRICAL FEEDBACK TO THE MR850 HUMIDIFIER. METHOD: THE TWO (2) COMPLAINT PROBES WERE RETURNED TO FISHER & PAYKEL HEALTHCARE'S REGIONAL OFFICE AND SERVICE CTR IN (B)(4). THE PROBES WERE TESTED FOR PERFORMANCE AND FUNCTIONALITY. RESULTS: TESTING REVEALED THAT THE FLOW THERMISTOR, WHICH TAKES MEASUREMENTS AT THE HUMIDIFICATION CHAMBER, OF THE FIRST PROBE WAS NOT WORKING CORRECTLY. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT # 090109. THE SECOND PROBE FUNCTIONED PROPERLY DURING TESTING. CONCLUSION: A FAULTY FLOW THERMISTOR WILL CAUSE THE HEATER BASE TO READ HIGHER THAN THE ACTUAL TEMPERATURE AND DROP AND THE HEAT SETTING. THE REDUCED TEMPERATURE SETTING WILL THEN RESULT IN THE CONDENSATION REPORTED BY THE HOSP. THE FAULTY PROBE HAS BEEN REPLACED. NO FAULT WAS FOUND WITH THE SECOND PROBE. (B)(4).

Description of Event or Problem · 1

A HOSP IN (B)(6) REPORTED THAT THEY ARE HAVING PROBLEMS WITH CONDENSATION. THE HOSP REPORTED THAT THEY BELIEVE THAT THE RAINOUT IS CAUSING INACCURATE MINUTE VOLUME READINGS ON THE VENTILATOR. THE HOSP REQUESTED THAT TWO (2) 900MR869 TEMPERATURE/FLOW PROBES BE TESTED FOR ACCURACY. NO PT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TEMPERATURE/FLOW PROBE BTT FISHER & PAYKEL HEALTHCARE, LTD. 900MR869 090109

Patients

Seq Age Sex Outcome Treatment
1 MR850 RESPIRATORY HUMIDIFIER| UNK THIRD-PARTY VENTILATOR| FISHER & PAYKEL HEALTHCARE RT200 BREATHING CIRCUIT