FDA Adverse Event Injury Summary report: N

BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE M

MDR report key: 16741093 · Received April 14, 2023

Report

Report Number
3005180920-2023-00229
Event Type
Injury
Date Received
April 14, 2023
Report Date
April 14, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809163
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 17 MARCH 2023. LOT 2011664: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-MARCH-2021. EXPIRATION DATE: 2026-02-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. NOTE: THE FIELD "DATE OF THE EVENT" IS EMPTY BECAUSE WE DON'T KNOW WHEN THE USE ERROR WAS DISCOVERED AND WHEN THE PATIENT UNDERWENT TO A REVISION SURGERY.

Description of Event or Problem · 0

DURING TOTAL HIP REPLACEMENT SURGERY ON (B)(6) 2021, WRONG SIZE OF THE HEAD, Ø32 INSTEAD OF Ø36 HAS BEEN IMPLANTED (LINER G MUST BE USED WITH HEAD Ø36). THE PATIENT UNDERWENT TO A REVISION SURGERY TO REPLACE THAT COMPONENT. THE REVISION SURGERY DATE IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1243731 BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE M HIP FEMORAL HEAD LZO MEDACTA INTERNATIONAL SA 01.29.205 2011664 07630030809163

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention