FDA Adverse Event Malfunction Summary report: N

630G INSULIN PUMP MMT-1715KR 630G

MDR report key: 16740919 · Received April 14, 2023

Report

Report Number
2032227-2023-188953
Event Type
Malfunction
Date Received
April 14, 2023
Date of Event
March 29, 2023
Report Date
May 19, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000367060
PMA / PMN Number
P150001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. THE PUMP PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST. THE PUMP WAS MONITORED FOR SEVERAL DAYS AND NO BLANK DISPLAY NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. THERE WERE NO UNEXPECTED PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE 29-MAR-2023 IN THE FORMATTED HISTORY FILE. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS/ALERTS WERE NOTED. INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 15:48:34.000, (B)(6) 2023 15:57:20.000, (B)(6) 2023 15:57:48.000 (B)(6) 2023 16:27:16.000, (B)(6) 2023 16:49:46.000, (B)(6) 2023 17:06:37.000 (B)(6) 2023 17:07:20.000, (B)(6) 2023 17:17:00.000, (B)(6) 2023 17:19:58.000 PUMP ERROR 68 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 17:22:27.000 PUMP ERROR 49 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 17:22:27.000 (B)(6) 2023 17:22:57.000 PUMP ERROR 23 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 17:18:03.000 (B)(6) 2023 17:23:08.000 POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 17:18:38.000 (B)(6) 2023 17:18:49.000 (B)(6) 2023 17:23:20.000 (B)(6) 2023 17:23:31.000 LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 14:46:00.000 (B)(6) 2023 16:18:00.000 (B)(6) 2023 16:35:00.000 FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2023 15:56:20.000, (B)(6) 2023 15:57:32.000, (B)(6) 2023 16:07:00.000 (B)(6) 2023 16:08:02.000, (B)(6) 2023 16:17:48.000, (B)(6) 2023 16:27:00.000 (B)(6) 2023 16:34:06.000, (B)(6) 2023 16:34:13.000, (B)(6) 2023 16:44:00.000 (B)(6) 2023 16:49:18.000, (B)(6) 2023 16:49:46.000, (B)(6) 2023 16:49:57.000 (B)(6) 2023 16:50:52.000, (B)(6) 2023 17:00:00.000 INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, LOW BATTERY ALERT AND FAILED BATTERY ALERT/BATTERY FAILED ALARM WERE EXPECTED SINCE THE BATTERY IN THE PUMP IS LOW ON POWER. THE CUSTOMER MAY HAVE USED A LOW POWER BATTERY. PUMP ERROR 23 ALARM AND POWER LOSS ALARM WERE EXPECTED SINCE THE PUMP RESET AFTER THE BATTERY WAS REMOVED FOR MORE THAN 10 MINUTES. PUMP ERROR 68 ALARM AND PUMP ERROR 49 ALARM WERE CONFIRMED DUE TO POWER LOSS. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND IT WAS VERIFIED THAT THE BATTERY TUBE POWER CONNECTOR IS PROPERLY CONNECTED TO J7/PCB1. NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY AND A SCRATCHED CASE. BLANK DISPLAY WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THE INSULIN PUMP NOT TURNING ON AFTER REPLACEMENT OF THE BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE BLANK DISPLAY. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INSULIN PUMP AND REVERT TO THE BACKUP PLAN AS PER HEALTHCARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1094929 630G INSULIN PUMP MMT-1715KR 630G AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715KR HG5GDAN 000000763000367060

Patients

Seq Age Sex Outcome Treatment
1 85 YR Female