FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1673856 · Received May 5, 2010

Report

Report Number
6000001-2010-00509
Event Type
Malfunction
Date Received
May 5, 2010
Date of Event
April 27, 2010
Report Date
April 28, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE FACILITY'S BIOMEDICAL TECHNICIAN (BT) CALLED BAXTER TECHNICAL SUPPORT CONCERNING A COLLEAGUE INFUSION PUMP EXPERIENCING FAILURE CODE 810:04. WITH THE HELP OF BAXTER TECHNICAL SUPPORT, THE BT CHECKED AND CONFIRMED THE AIR IN LINE (AIL) PRINTED CIRCUIT BOARD (PCB) WAS OUT OF CALIBRATION. THE BT RE-CALIBRATED THE AIL PCB TO CORRECT THE REPORTED CONDITION. THE FACILITY WILL NOT BE SENDING THIS PUMP TO BAXTER FOR DEVICE EVALUATION OR REPAIR. HOWEVER, SHOULD THIS PUMP BE RECEIVED BY BAXTER FOR EVALUATION, OR IF ANY ADDITIONAL INFORMATION ABOUT THIS EVENT BECOMES AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). ISSUES CONCERNING AIR IN LINE PRINTED CIRCUIT BOARD FAILURES ARE CURRENTLY BEING INVESTIGATED UNDER (B)(4).

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:04, DUE TO THE PUMP'S AIR IN LINE PRINTED CIRCUIT BOARD BEING OUT OF CALIBRATION. THIS EVENT WAS REPORTED TO HAVE OCCURRED DURING SET PRIMING IN THE NEONATAL INTENSIVE CARE UNIT. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 CLEARLINK BURETROL SET ((B)(4))