GN RESOUND
Report
- Report Number
- 3005650109-2023-00046
- Event Type
- Malfunction
- Date Received
- April 13, 2023
- Date of Event
- March 20, 2023
- Report Date
- April 13, 2023
- Manufacturer
- GN HEARING A/S
- Product Code
- OSM
- UDI-DI
- 05708296147113
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MANUFACTURER'S REF# (B)(4). TECHNICAL INVESTIGATION CONCLUDED: A DEVICE HISTORY RECORD REVIEW HAVE BEEN COMPLETED. NO DEVIATION OR CHANGES WERE FOUND DURING MANUFACTURING OF THE DEVICE. TRENDED CASE CONCLUDED: THE CONNECTOR IS BROKEN, PROBABLY BECAUSE THE USER HAS (NOT INTENTIONALLY) MISUSED IT, WHILE INSERTING THE CABLE. IF THE CABLE HAS BEEN SLIGHTLY OUT OF SPECIFICATION THAT COULD HAVE HAD A NEGATIVE EFFECT ON THIS. THIS HAS CAUSED A LOW OHMIC CONNECTION IN THE CABLE CONNECTOR - ALMOST A SHORT CIRCUIT - THAT WILL EMIT HEAT WHEN THE CABLE IS INSERTED INTO THE WALL CHARGER, CAUSING UP TO THE RATED 5 W OF THE WALL CHARGER TO BE DISSIPATED INSIDE THE CONNECTOR. MICRO USB IS A VERY WELL KNOWN AND USED STANDARD BUT CAN MECHANICALLY BREAK DOWN. THE CLINICAL INVESTIGATION CONCLUDED: A DAMAGED POWER SUPPLY OR A POWER SUPPLY OF VERY POOR QUALITY COULD LEAD TO BURN INJURIES AND IN RARE CONDITIONS THERE IS A SMALL RISK THAT A WRONG CHARGING OR DISCHARGING OR A BATTERY THERMAL RUNWAY CAN LEAD TO HIGH TEMPERATURES OR FIRE ACCIDENTS. THE CHARGERS HAVE BEEN TESTED ACCORDING TO APPLICABLE SAFETY STANDARDS. THESE RISKS ARE REDUCED AS FAR AS POSSIBLE AND THE PROBABILITY OF OCCURRENCE OF THESE EVENTS IS CONSIDERED VERY LOW. THERE ARE NO NEW CLINICAL ASPECTS (E.G. UNFORESEEN USE OR CLINICAL CONDITIONS) DISCOVERED FROM THIS EVENT. THEREFORE, IT IS CONCLUDED THAT THE CURRENT CLINICAL EVALUATION AND RISK/BENEFIT CONCLUSION HEREIN ARE SUFFICIENT. RISK REGISTER CONCLUSION REFERENCE: THE RESIDUAL RISK AFTER THE RISK CONTROL MEASURE(S) IS ACCEPTABLE. NO NEW HAZARDOUS SITUATIONS REVEALED AND NO NEED TO UPDATE THE RISK REGISTER. MANUFACTURER'S INVESTIGATION IS FINAL. THIS IS A COMBINED (INITIAL AND FINAL) REPORT.
ESTIMATED DATE OF INCIDENT (B)(6) 2023. IT WAS REPORTED- CHARGER MELTED CHARGING PORT (OVER HEATING). BASED ON DESCRIPTION IT IS WHERE THE CABLE PLUGS INTO THE CHARGER (BY THE MINI USB ADAPTER CORD) THE END USER WENT TO PULL THE, THE CHARGING BOX OFF THE PLUG AND PLUG IT IN AND, SHE REALIZED IT WAS OBVIOUSLY SCORCHED AND, THE CHARGER ITSELF IN THE BACK IS MULTI PLASTIC AND THE PORTS THAT GOES INTO THE SEAPORT IS TOTALLY MELTED AND DISFIGURED. NO HARM DONE TO THE PATIENT OR TO THEIR HOME. ORIGINAL EQUIPMENT WAS USED. THE CHARGER IS OUT OF WARRANTY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1456847 | GN RESOUND | HEARING AID, , AIR CONDUCTION WITH WIRELESS TECHNOLOGY | OSM | GN HEARING A/S | 20472100 | 05708296147113 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female |