FDA Adverse Event Other Summary report: N

HT50 VENTILATOR

MDR report key: 1673414 · Received April 26, 2010

Report

Report Number
3003135857-2010-00006
Event Type
Other
Date Received
April 26, 2010
Date of Event
March 25, 2010
Report Date
March 29, 2010
Manufacturer
FLIGHT MEDICAL
Product Code
CBK
PMA / PMN Number
K082724
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: THE REPORTED PROBLEM WITH ALL BUTTONS UNRESPONSIVE WAS DUPLICATED; HOWEVER, THE VENTILATOR SHUTDOWN WAS NOT CONFIRMED. A VISUAL INSPECTION OF THE RETURNED VENTILATOR DID NOT FIND ANYTHING UNUSUAL. WHEN THE RETURNED VENTILATOR WAS FIRST POWERED ON, IT DID NOT TURN ON BECAUSE THE BATTERY WAS COMPLETELY DEPLETED: NO UNIT SHUTDOWN WAS OBSERVED. THE BATTERY WAS CHARGED AND THE VENTILATOR WAS TURNED ON; HOWEVER, THE UNIT SHUTDOWN WAS STILL NOT DUPLICATED. WHEN THE VENTILATOR WAS CONNECTED TO AC POWER, IT IMMEDIATELY STARTED BEEPING. AIR WAY PRESSURE METER NEEDLE FLUCTUATED, THE DISPLAY BECAME BLANK AND ALL BUTTONS WERE UNRESPONSIVE. THE SUSPECTED CONTROL PANEL BOARD IN THE CUSTOMER'S VENTILATOR WAS REMOVED AND A KNOWN GOOD CONTROL PANEL BOARD WAS INSTALLED. RESULT: THE SAME PROBLEM OCCURRED. THE SUSPECTED MAIN BOARD IN THE CUSTOMER'S VENTILATOR WAS REMOVED AND A KNOWN GOOD MAIN BOARD WAS INSTALLED. RESULT: THE VENTILATOR FUNCTIONED NORMALLY AND THE BUTTONS WERE RESPONSIVE. THE VENTILATOR REGAINED A NORMAL FUNCTIONALITY AFTER MAIN BOARD AND CONTROL PANEL BOARD WERE REPLACED. THE FAULTY PARTS WILL BE FORWARDED TO THE MANUFACTURER FOR FURTHER ANALYSIS. CONTROL PANEL BOARD.

Description of Event or Problem · 1

REPORTEDLY, THE VENTILATOR SHUTDOWN DURING USE ON A PT. THE VENTILATOR KEPT ALARMING AFTER IT SHUTDOWN AND ALL BUTTONS WERE UNRESPONSIVE. THE VENTILATOR FINALLY RAN DOWN DUE TO THE DRAINED BATTERY POWER. THE VENTILATOR DID NOT GIVE A PROPER ALARM PRIOR TO SHUTDOWN. THE PT WAS MANUALLY VENTILATED AND THEN SWITCHED TO THE BACKUP VENTILATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HT50 VENTILATOR CBK FLIGHT MEDICAL HT50-H NA

Patients

Seq Age Sex Outcome Treatment
1 27 YR Required Intervention