FDA Adverse Event Injury Summary report: N

CARE TOUCH

MDR report key: 16728937 · Received April 12, 2023

Report

Report Number
3016774562-2023-00745
Event Type
Injury
Date Received
April 12, 2023
Report Date
October 4, 2023
Manufacturer
TIANJIN EMPECS MEDICAL DEVICE CO., LTD.
Product Code
NBW
Adverse Event
Yes
Report Source
Distributor report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Description of Event or Problem · 0

CUSTOMER COMPLAINT: LIMITED DATA AVAILABLE. I PURCHASED A CARETOUCH A COUPLE WEEKS AGO AND ORDERED A CONTROL SOLUTION. THE CONTROL SOLUTION ARRIVED TODAY AND IT REPEATEDLY FAILS QA FOR THE LEVEL 2 SOLUTION. THE RANGE ON THE STRIPS IS (CODE A01 LOT AWD07H1A1 EXP 2024-04-06) L2: 68-92 L3 216-293,BUT 4 L2 TESTS YIELDED RESULTS OF 64, 67, 66, 66. THE LEVEL 3 SAMPLE REPORTED AS 221 AND 234. I PURCHASED THROUGH AMAZON AND THEY PROVIDED A REFUND FOR THE ISSUE, BUT I RATHER HAVE THE METER WORKING AND REVERSE THE REFUND FROM THE SELLER AND I WANTED TO GIVE YOU AN OPPORTUNITY TO RESOLVE THE ISSUE. THE CONTROL SOLUTION IS LOT 1AR2A091AR3A07, IT EXPIRES IN 231114.

Description of Event or Problem · 0

CUSTOMER COMPLAINT - LIMITED DATA AVAILABLE . THE CUSTOMER STATED THAT THEY WERE RECIEVING INACCUATE READINGS WITH THEIR BLOOD GLUCOSE MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1306109 CARE TOUCH BLOOD GLUCOSE MONITOR KIT NBW TIANJIN EMPECS MEDICAL DEVICE CO., LTD. CT210 1AR2A091AR3A07, AWD07H1A1

Patients

Seq Age Sex Outcome Treatment
1 Male