CARE TOUCH
Report
- Report Number
- 3016774562-2023-00745
- Event Type
- Injury
- Date Received
- April 12, 2023
- Report Date
- October 4, 2023
- Manufacturer
- TIANJIN EMPECS MEDICAL DEVICE CO., LTD.
- Product Code
- NBW
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CUSTOMER COMPLAINT: LIMITED DATA AVAILABLE. I PURCHASED A CARETOUCH A COUPLE WEEKS AGO AND ORDERED A CONTROL SOLUTION. THE CONTROL SOLUTION ARRIVED TODAY AND IT REPEATEDLY FAILS QA FOR THE LEVEL 2 SOLUTION. THE RANGE ON THE STRIPS IS (CODE A01 LOT AWD07H1A1 EXP 2024-04-06) L2: 68-92 L3 216-293,BUT 4 L2 TESTS YIELDED RESULTS OF 64, 67, 66, 66. THE LEVEL 3 SAMPLE REPORTED AS 221 AND 234. I PURCHASED THROUGH AMAZON AND THEY PROVIDED A REFUND FOR THE ISSUE, BUT I RATHER HAVE THE METER WORKING AND REVERSE THE REFUND FROM THE SELLER AND I WANTED TO GIVE YOU AN OPPORTUNITY TO RESOLVE THE ISSUE. THE CONTROL SOLUTION IS LOT 1AR2A091AR3A07, IT EXPIRES IN 231114.
CUSTOMER COMPLAINT - LIMITED DATA AVAILABLE . THE CUSTOMER STATED THAT THEY WERE RECIEVING INACCUATE READINGS WITH THEIR BLOOD GLUCOSE MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1306109 | CARE TOUCH | BLOOD GLUCOSE MONITOR KIT | NBW | TIANJIN EMPECS MEDICAL DEVICE CO., LTD. | CT210 | 1AR2A091AR3A07, AWD07H1A1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |