FDA Adverse Event Malfunction Summary report: N

COMPLETE MULTI-PURPOSE EASY RUB FORMULA

MDR report key: 1672874 · Received April 29, 2010

Report

Report Number
2020664-2010-00011
Event Type
Malfunction
Date Received
April 29, 2010
Date of Event
November 25, 2009
Report Date
November 25, 2009
Manufacturer
ABBOTT MEDICAL OPTICS, INC.
Product Code
LPN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MFG RECORDS FOR THE REPORTED LOT (AD00696) WAS PERFORMED; NO DEVIATIONS WERE NOTED AND PRODUCT MET ALL SPECS. CHEMISTRY EVAL RESULTS OF THE COMPLAINT UNIT AND A RETAINED UNIT FROM THE REPORTED LOT WERE WITHIN SPECS. MICROBIOLOGY EVAL OF A RETAINED UNIT FROM THE REPORTED LOT SHOWED THE SOLUTION TO BE STERILE. THE COMPLAINT UNIT WAS RETURNED FOR MICROBIOLOGY TESTING. BUT AS THE COMPLAINT UNIT WAS ALREADY OPENED, MICROBIOLOGY RESULTS SHOW THAT THE SAMPLE WAS NOT STERILE. USER ERROR MAY HAVE CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 1

OUR OFFICE IN (B) (4) REC'D A REPORT FROM THE MOTHER OF A MALE PT THAT HER SON EXPERIENCED AN "INFECTION" AFTER USE OF COMPLETE MULTI-PURPOSE SOLUTION EASY RUB FORMULA (SEE RELATED COMPLAINT, MDR #2020664-2009-00088). THERE IS NO INDICATION THAT HE REC'D ANY MEDICAL ATTENTION, AND HIS SYMPTOMS HAVE RESOLVED. SHE DESCRIBED HIS LENS CARE REGIMEN AS ALWAYS USING FRESH SOLUTION AND RINSING HIS LENS CASE WITH COMPLETE MPS. HE DID NOT RUB HIS CONTACTS BUT SHE REPORTED THAT HE WOULD SHAKE THE LENS CASE INSTEAD. PRIOR TO THE INFECTION, THE PT WAS USING MONTHLY DISPOSABLE CONTACT LENSES; HE IS NOW USING DAILY DISPOSABLE CONTACTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPLETE MULTI-PURPOSE EASY RUB FORMULA LPN ABBOTT MEDICAL OPTICS, INC. NA AD00696

Patients

Seq Age Sex Outcome Treatment
1