LIBERTY CYCLER
Report
- Report Number
- 2937457-2010-00005
- Event Type
- Other
- Date Received
- April 29, 2010
- Date of Event
- March 31, 2010
- Report Date
- March 31, 2010
- Manufacturer
- FRESENIUS MEDICAL CARE - NORTH AMERICA
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4) - CYCLER ADVANCED FROM DRAIN 0 TO FILL 1 AFTER 3 MINUTES. THE CYCLER MONITORS THE AMOUNT OF SOLUTION DELIVERED TO OR DRAINED FROM THE PT DURING TREATMENT. MANUAL EXCHANGES COULD NOT BE MONITORED BY THE CYCLER. IF THE LAST FILL PROGRAMMED WAS 0, THE CYCLER WILL ADVANCE FROM DRAIN 0 TO FILL 1 WHEN THE FLOW OF SOLUTION STOPS SINCE NO SOLUTION IS EXPECTED TO BE IN THE PERITONEAL CAVITY WHEN TREATMENT IS STARTED. THE OVERFILL OCCURRED IN THIS EVENT BECAUSE THE PT PERFORMED A MANUAL FILL DURING THE DAY WITH NO LAST FILL PROGRAMMED. (B)(4).
A HOME PERITONEAL DIALYSIS PT'S WIFE REPORTED THAT THE CYCLER SKIPPED THE DRAIN 0 CYCLE. THE RECORDS SHOWED THAT THE PT DRAINED A TOTAL 358 ML. IN 3 MINUTES IN DRAIN 0 BEFORE SWITCHING TO FILL 1. THE PT C/O DIFFICULTY BREATHING, BECAME VERY LIGHT-HEADED AND THOUGHT HE WAS GOING TO PASS OUT. THE WIFE REALIZED THAT THE PT HAD NOT FULLY DRAINED THE SOLUTION THAT THE PT WAS FILLED MANUALLY DURING THE DAY, WHEN THE CYCLER SWITCHED TO THE NEXT FILL. SHE DRAINED HIM MANUALLY AND WAS ABLE TO DRAIN 3,500 ML IN THE FIRST DRAIN BAG AND 2,200 ML IN THE SECOND BAG. THE PT FELT BETTER AFTER DRAINING WITH THE FIRST BAG. THE PT'S PRESCRIPTION IS FOR FILL VOLUMES OF 2,500 ML. WITH NO LAST FILL. THE PT IS EMPTY WHEN HE DISCONNECTS FROM THE CYCLER IN THE MORNING. HE PERFORMS A MANUAL FILL OF 2,500 ML. DURING THE DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIBERTY CYCLER | PERITONEAL DIALYSIS CYCLER | FKX | FRESENIUS MEDICAL CARE - NORTH AMERICA | LIBERTY | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |