FDA Adverse Event Malfunction Summary report: N

DIALOG+®

MDR report key: 16722966 · Received April 12, 2023

Report

Report Number
3002879653-2023-00014
Event Type
Malfunction
Date Received
April 12, 2023
Date of Event
November 1, 2023
Report Date
April 12, 2023
Manufacturer
B. BRAUN AVITUM AG
Product Code
KDI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL INC. WE RECEIVED A REPORT OF 11 PATIENTS COMPLAINING OF EXTREME THIRST AND/OR FEELING UNWELL DURING DIALYSIS THERAPY. TWO PATIENTS EXPERIENCED DIFFICULTY OF BREATHING. THESE PATIENTS CAME FOR ADDITIONAL DIALYSIS TREATMENT. NO FURTHER MEDICAL INTERVENTION WAS NEEDED. THE DIALYSIS THERAPIES TOOK PLACE ON FOUR DIFFERENT MACHINES WITH SERIAL NUMBERS (B)(4). ALL OF THE AFFECTED DEVICES WERE ISOLATED AND INSPECTED BY A LOCAL B. BRAUN'S BIOMEDICAL TECHNICIAN. INSPECTION REVEALED THAT THERE WAS A DEVIATION IN THE FINAL CONDUCTIVITY OF APPROXIMATELY 2 MS/CM IN THE MACHINES CONCERNED. THE CONSEQUENCE OF THIS DEVIATION IS THAT THE PATIENTS WERE TREATED WITH A HIGHER SODIUM CONCENTRATION IN THE DIALYSIS FLUID THAN INTENDED, THUS EXPLAINING THEIR THIRST. ANALYSIS OF THE SERVICE REPORTS OF THE AFFECTED MACHINES SHOWED THAT ONE EXTERNAL CONTRACT TECHNICIAN HAD PREVIOUSLY CALIBRATED THE TEMPERATURE AND CONDUCTIVITY SENSORS ON THESE MACHINES. IN ORDER TO FIND THE ROOT CAUSE OF THE DEVIATION, THE TECHNICIAN'S MEASURING EQUIPMENT WAS SENT TO A TESTING LABORATORY FOR VERIFICATION. THE LABORATORY'S TEST SHOWED NO DEVIATION OF THE MEASURING DEVICE WHICH COULD RESULT IN A CONDUCTIVITY DEVIATION OF APPROXIMATELY 2 MS/CM. FOR FURTHER ROOT CAUSE ANALYSIS A TECHNICIAN OF THE GLOBAL TECHNICAL SERVICE OF B. BRAUN HEADQUARTER, MELSUNGEN, GERMANY, WAS SENT TO AUSTRALIA TO VERIFY ALL MACHINES IN THE RESPECTIVE DIALYSIS CENTER. FOR EACH MACHINE WHICH WAS PREVIOUSLY SERVICED BY THE EXTERNAL CONTRACT TECHNICIAN A CHECK AND RECALIBRATION WAS PERFORMED. THE CHECK OF THE AFFECTED MACHINES BY THE GLOBAL TECHNICAL SERVICE OF B. BRAUN HEADQUARTER CONFIRMED THAT THE EXTERNAL CONTRACT MANUFACTURER HAD SET A WRONG VALUE WHEN CALIBRATING THE CONDUCTIVITY SENSORS. TO AVOID A RECURRENCE, THE EXTERNAL CONTRACT TECHNICIAN WAS RETRAINED ON SITE BY THE HEADQUARTERS' GLOBAL TECHNICAL SERVICE TECHNICIAN. GENERALLY, EACH DIALYSIS TECHNICIAN SERVICING B. BRAUN MACHINES IS CERTIFIED BY TRAINING. CALIBRATION OF CONDUCTIVITY AND TEMPERATURE SENSORS IS DESCRIBED AND EXPLAINED IN DETAIL IN THE TECHNICAL SERVICE MANUAL AND WITHIN THE TECHNICAL TRAINING COURSE FOR CERTIFICATION. SINCE CHECK AND RECALIBRATION BY THE HEADQUARTERS' GLOBAL TECHNICAL SERVICE TECHNICIAN NO FURTHER ABNORMALITIES OCCURRED. (B)(4).

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: PATIENTS HAVE BEEN COMPLAINING OF BEING EXTREMELY THIRSTY DURING DIALYSIS TREATMENT AND SOME FEEL UNWELL. PATIENTS DRINKING MORE FLUID THAN USUAL BECAUSE OF EXTREME THIRST. TWO PATIENTS EXPERIENCED DIFFICULTY IN BREATHING HENCE REQUIRED THEM TO COME FOR AN EXTRA SESSION OF DIALYSIS. THIS REPORT IS FOR EVENT #8.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2253194 DIALOG+® KDI-DIALYZER, HIGH PERMEABILIT KDI B. BRAUN AVITUM AG

Patients

Seq Age Sex Outcome Treatment
1 Unknown