BINAXNOW COVID-19 AG SELF TEST 2CT
Report
- Report Number
- 1221359-2023-00613
- Event Type
- Malfunction
- Date Received
- April 11, 2023
- Date of Event
- April 1, 2023
- Report Date
- April 11, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
THE EVENTS DATE IS AN ESTIMATE. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210543 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210543 AND DEVICE PART NUMBER 195-430WL / LOT 207460. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210543 SHOWED THAT THE COMPLAINT RATE IS 0.00425%. ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE E, HOWEVER IT COULD HAVE POSSIBLY BEEN RELATED TO THE SPECIFIC PATIENT SAMPLE. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. ABBOTT DIAGNOSTICS SCARBOROUGH WILL CONTINUE TO MONITOR AND TREND FOR THIS REPORTED ISSUE.
THE CONSUMER REPORTED NINE (9) FALSE NEGATIVE RESULTS WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6)2023 ON NASAL SAMPLE. THE CONSUMER RECEIVED NINE (9) NEGATIVE RESULTS WITH BINAXNOW COVID-19 AG SELF-TEST, AND A POSITIVE RESULT AT THE DOCTOR CLINIC. THE CONSUMER HAD MUSCLE ACHES AND HEADACHE AND SORE THROAT. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF NINE (9) TESTS, IN REGARD TO LOT NUMBER (210543) QUANTITY (5). NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1720964 | BINAXNOW COVID-19 AG SELF TEST 2CT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 210543 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Female |