FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE¿ PRO PEN NEEDLE

MDR report key: 16713485 · Received April 11, 2023

Report

Report Number
9616656-2023-00347
Event Type
Malfunction
Date Received
April 11, 2023
Date of Event
March 20, 2023
Report Date
April 27, 2023
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: ONE OPEN 32G X 4MM PEN NEEDLE SAMPLE AND THREE PHOTOS WERE RETURNED FROM LOT. NO. 2110717, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLE AND PHOTOS AND BLACK MARKS WERE OBSERVED ON THE COVER. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. BLACK MARKS OCCURRED DUE TO THE MOULDING PRESS STOPPING WHILST HEATS REMAINED ON THE INJECTION UNIT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT INK-LIKE FOREIGN MATTER WAS FOUND ON THE NON-PATIENT END OF THE BD MICRO-FINE¿ PRO PEN NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE: "ACCORDING TO THE USER'S VERBATIM REPORT, WHEN OPENING THE TEAR DROP LABEL, THERE WAS SOMETHING LIKE A INK ON THE NEEDLE NPE."

Description of Event or Problem · 0

IT WAS REPORTED THAT INK-LIKE FOREIGN MATTER WAS FOUND ON THE NON-PATIENT END OF THE BD MICRO-FINE¿ PRO PEN NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE: "ACCORDING TO THE USER'S VERBATIM REPORT, WHEN OPENING THE TEAR DROP LABEL, THERE WAS SOMETHING LIKE A INK ON THE NEEDLE NPE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402257 BD MICRO-FINE¿ PRO PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 2110717

Patients

Seq Age Sex Outcome Treatment
1 Unknown