FDA Adverse Event Malfunction Summary report: N

FREE STANDING INFANT WARMER

MDR report key: 167124 · Received May 7, 1998

Report

Report Number
2510954-1998-00004
Event Type
Malfunction
Date Received
May 7, 1998
Date of Event
April 7, 1998
Report Date
April 7, 1998
Manufacturer
HILL-ROM AIR-SHIELDS
Product Code
FMT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
SERVICE PERSONNEL

Narratives

Description of Event or Problem · 1

A HILL-ROM AIR-SHIELDS SERVICE ENGINEER NOTICED A BROKEN WELD ON THE LOWER POST OF THE RADIANT WARMER. THE WELDING FAULT WAS DETECTED DURING SETUP. THE BROKEN WELD WAS LOCATED AT THE BOTTOM OF THE POST, WHERE THE POST MOUNTS TO THE BASE, IN ORDER TO SUPPORT THE WARMER MODULE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREE STANDING INFANT WARMER INFANT RADIANT WARMER FMT HILL-ROM AIR-SHIELDS 7830 NA

Patients

Seq Age Sex Outcome Treatment
1 NA