FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN

MDR report key: 16710708 · Received April 10, 2023

Report

Report Number
2243072-2023-00585
Event Type
Malfunction
Date Received
April 10, 2023
Date of Event
March 21, 2023
Report Date
May 15, 2023
Manufacturer
BECTON DICKINSON
Product Code
JKA
UDI-DI
50382903687744
PMA / PMN Number
K101502
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IN THIS MDR, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AS SEKISUI IS AN OEM MANUFACTURING SITE. D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. H.4. DEVICE MANUFACTURE DATE: UNKNOWN.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES, D9: RETURNED TO MANUFACTURER ON: 20223-04-10. H.6. INVESTIGATION SUMMARY: BD RECEIVED 100 SAMPLES AND 1 PHOTO FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WITH THE INCIDENT LOT WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER CENTRIFUGATION WITH BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN THERE WAS POOR BARRIER SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SM 368774 BARRIER FORMATION FAILED AFTER CENTRIFUGATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER CENTRIFUGATION WITH BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN THERE WAS POOR BARRIER SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SM 368774 BARRIER FORMATION FAILED AFTER CENTRIFUGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1623951 BD VACUTAINER® RAPID SERUM TUBE (RST) BLOOD COLLECTION TUBES THROMBIN BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON DICKINSON 368774 230201 50382903687744

Patients

Seq Age Sex Outcome Treatment
1 Unknown