FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE

MDR report key: 16709810 · Received April 10, 2023

Report

Report Number
3003152976-2023-00124
Event Type
Malfunction
Date Received
April 10, 2023
Date of Event
March 21, 2023
Report Date
April 27, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE SAMPLE RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL EVALUATION, BARREL IS DAMAGED, WHICH LIKELY CAUSED THE LEAKAGE PAST STOPPER. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2212074, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. POSSIBLE ROOT CAUSE IS ASSOCIATED WITH MISALIGNMENT IN THE ASSEMBLY PROCESS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: WE HAVE A BATCH OF LEAKING SYRINGES REF 300865. LOT 2212074. WE HAVE KEPT ONE SYRINGE FOR EXPERTISE.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: WE HAVE A BATCH OF LEAKING SYRINGES REF 300865. LOT 2212074. WE HAVE KEPT ONE SYRINGE FOR EXPERTISE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1798777 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2212074

Patients

Seq Age Sex Outcome Treatment
1 Unknown