FDA Adverse Event
Malfunction
Summary report: N
GOODMAN CO. LTD
MDR report key: 1670957
·
Received April 26, 2010
Report
- Report Number
- 2243801-2010-00005
- Event Type
- Malfunction
- Date Received
- April 26, 2010
- Date of Event
- February 12, 2010
- Report Date
- April 20, 2010
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- DYB
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE ACTUAL DEVICE IN THE INCIDENT WAS RETURNED TO THE MANUFACTURER TO BE EVALUATED. CONTINUITY TESTING WAS PERFORMED PER SPECIFICATION ON THE RETURNED CATHETER SAMPLE. THE TESTING WAS CONDUCTED WITH VIGOROUS MANIPULATION TO THE CATHETER BODY, MANIFOLD AND LEADS (EXCESSIVE BENDING, SQUEEZING, ETC.) AND NO ABNORMALITY WAS FOUND. THE PROXIMAL LINE AND DISTAL LINE HAD PERFECT CONTINUITY AND THERE WAS NO SHORT CIRCUIT IN THE CATHETER ELECTRICAL SYSTEM. THERE WAS NO LOSS IN CONTINUITY THAT COULD RESULT IN PACING FAILURE. THERE ARE NO OTHER REPORTS OF THIS NATURE AGAINST THE REPORTED LOT NUMBER.
Description of Event or Problem · 1
REPORTS PACING FAILURE WHILE IN USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GOODMAN CO. LTD | PACING CATHETER | DYB | B. BRAUN MEDICAL, INC. | NA | 61045800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |