PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Report
- Report Number
- 2024168-2023-03639
- Event Type
- Injury
- Date Received
- April 10, 2023
- Date of Event
- March 17, 2023
- Report Date
- June 23, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PROSTYLE DEVICE REFERENCED IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.
THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED NEEDLE TO CUFF MISS WAS NOT CONFIRMED AS NOT ALL COMPONENTS WERE RETURNED FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THE FOOT SEPARATION WAS LIKELY DUE TO HANDLING OR MANIPULATION OF THE DEVICE POST PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH TWO PROSTYLE DEVICES AFTER A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE USING A 6F SHEATH. REPORTEDLY, THE SUTURE WAS NOT PRESENT WHEN THE PLUNGER WAS PULLED BACK FOR BOTH DEVICES. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED:A POSTERIOR FOOT BREAK WAS FOUND DURING EVALUATION OF THE RETURNED DEVICE (CN-161379). THE LOCATION OF THE DETACHED FOOT IS UNKNOWN. FOLLOW-UP WITH THE ACCOUNT CONFIRMED THAT THE FOOT SEPARATION DID NOT OCCUR DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1754019 | PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 2110741 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |