BD PRECISIONGLIDE NEEDLE
Report
- Report Number
- 1911916-2023-00222
- Event Type
- Malfunction
- Date Received
- April 8, 2023
- Date of Event
- March 22, 2023
- Report Date
- April 20, 2023
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMI
- UDI-DI
- 30382903051091
- PMA / PMN Number
- K021475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INITIAL REPORTER PHONE #: (B)(6) B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305109 AND LOT NUMBER 8250590. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. H3 OTHER TEXT : SEE H10
IT WAS REPORTED THAT THE BD PRECISIONGLIDE NEEDLE EXPERIENCED LEAKAGE AT THE BASE OF THE NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKING AT THE BASE OF THE NEEDLE, NOT AT THE JUNCTION.
IT WAS REPORTED THAT THE BD PRECISIONGLIDE NEEDLE EXPERIENCED LEAKAGE AT THE BASE OF THE NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKING AT THE BASE OF THE NEEDLE, NOT AT THE JUNCTION
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1753854 | BD PRECISIONGLIDE NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 305109 | 8250590 | 30382903051091 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |