FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 16705140 · Received April 8, 2023

Report

Report Number
1213809-2023-00335
Event Type
Malfunction
Date Received
April 8, 2023
Date of Event
February 16, 2023
Report Date
May 3, 2023
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY FOURTEEN SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE REPORTED ISSUE WAS NOT CONFIRMED UPON INSPECTION OF THE SAMPLES. NONE OF THE SAMPLES SHOWED ANY SIGNS OF THE REPORTED DEFECT. SINCE THE REPORTED DEFECT WAS NOT CONFIRMED A MANUFACTURING ROOT CAUSE COULD NOT BE DETERMINED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2030421. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD OIL STAINS ON/INSIDE IT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING OIL STAINS, UNCLEAR SCALE LINES, AND SURFACE SCRATCHES UPON THEIR INSPECTION... NO PATIENT HARM INVOLVED."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD OIL STAINS ON/INSIDE IT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING OIL STAINS, UNCLEAR SCALE LINES, AND SURFACE SCRATCHES UPON THEIR INSPECTION... NO PATIENT HARM INVOLVED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1729893 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 2030421

Patients

Seq Age Sex Outcome Treatment
1 Unknown