FDA Adverse Event Malfunction Summary report: N

BD ALARIS SMARTSITE NEEDLE-FREE VALVE

MDR report key: 16694965 · Received April 6, 2023

Report

Report Number
9616066-2023-00643
Event Type
Malfunction
Date Received
April 6, 2023
Date of Event
March 7, 2023
Report Date
March 21, 2023
Manufacturer
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
Product Code
FPA
UDI-DI
10885403238345
PMA / PMN Number
K061285
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

2000E CHINA 510K: THIS IS AN INTERNATIONAL CODE - THE MODEL#/CATALOG# IDENTIFIED IN SECTION IS A CAREFUSION PRODUCT WHICH IS SAME OR SIMILAR TO A DEVICE THAT IS APPROVED FOR SALE DOMESTICALLY. THE DOMESTIC SIMILAR LIST NUMBER IS 2000E. THE 510K NUMBER PROVIDED IN SECTION G5 IS FOR THE DOMESTIC SIMILAR PRODUCT: K061285. INVESTIGATION SUMMARY: A 2000E CHINA PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION; HOWEVER, THE CUSTOMER INDICATED THE COMPLAINT SAMPLE WAS FROM LOT 22045361. THE FEEDBACK PROVIDED BY THE CUSTOMER SUGGESTS OCCLUSION WAS DETECTED DURING USE OF THE SMARTSITE DEVICE. THE DETAILS OF THIS FEEDBACK WERE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. A REVIEW OF THE PRODUCTION RECORDS FOR LOT 22045361 DID NOT IDENTIFY ANY IN-PROCESS TESTING FAILURES OR QUALITY DEVIATIONS WHICH MAY HAVE RESULTED IN A REPORT OF THIS NATURE. THE ROOT CAUSE OF THE CUSTOMER¿S EXPERIENCE COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION. IN THIS INSTANCE, WITHOUT A SAMPLE, IT IS NOT POSSIBLE TO DETERMINE WHETHER A MANUFACTURING DEFECT COULD HAVE CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S EXPERIENCE. PREVIOUS COMPLAINTS FOR OCCLUSIONS HAVE BEEN RELATED TO FEATURES ON THE SURFACE OF THE MALE LUER OF THE CONNECTING PRODUCTS. THESE FEATURES INCLUDE FLASH OR A RAISED EDGE TO THE TIP OF THE MALE LUER WHICH HAVE PREVIOUSLY BEEN SHOWN TO INTERMITTENTLY LEAD TO RESTRICTED FLOW DUE TO THEM PINCHING THE BLUE PISTON OF THE SMARTSITE® AND NOT ALLOWING IT TO OPEN. THIS CAN SOMETIMES BE RESOLVED BY DISCONNECTING AND REATTACHING THE SAME LUER CONNECTION WHICH MAY REPOSITION THE LUER AGAINST THE PISTON AND IMPROVE THE FLOW, OR ALTERNATIVELY BY CHANGING THE CONNECTING MALE LUER. A REVIEW OF THE CUSTOMER FEEDBACK DATABASE INDICATES THAT THIS IS A RARE OCCURRENCE WITH A SMALL NUMBER OF SIMILAR REPORTS AGAINST THE SMARTSITE COMPONENT IN THE PAST 12 MONTHS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ALARIS SMARTSITE NEEDLE-FREE VALVE EXPERIENCED FLOW ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN USED FOR THE PATIENT, IT IS BLOCKED, RESULTING IN POOR FLUID REHYDRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1753221 BD ALARIS SMARTSITE NEEDLE-FREE VALVE INTRAVASCULAR ADMINISTRATION SET FPA SISTEMAS MEDICOS ALARIS, S.A. DE C.V. 22045361 10885403238345

Patients

Seq Age Sex Outcome Treatment
1 Unknown