ENTERALITE INFINITY ENTERAL FEEDING PUMP
Report
- Report Number
- 1722139-2010-00053
- Event Type
- Other
- Date Received
- April 23, 2010
- Date of Event
- April 5, 2010
- Report Date
- April 22, 2010
- Manufacturer
- MOOG MEDICAL DEVICES GROUP
- Product Code
- LZH
- PMA / PMN Number
- K031199
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
METHOD: ACTUAL DEVICE WAS EVALUATED. RESULTS: THE EVAL INCLUDED PERFORMANCE TESTING (ACCURACY, VERIFICATION OF AIR/NO FOOD ALARM, ETC). MULTIPLE FORMULAS AND SETTINGS WERE USED TO RECREATE THE FAILURE EXPERIENCED BY THE CUSTOMER. THE PUMP PERFORMED ACCORDING TO SPEC DURING EACH RUN (ALARMED AND SHUT OFF AFTER FOOD WAS GONE). CONCLUSION: THE ALLEGED COMPLAINT/DEFECT COULD NOT BE DUPLICATED. THE PUMP PERFORMED ACCORDING TO SPEC. THE DISPOSABLE SET USED AT THE TIME OF THE FAILURE WAS NOT RETURNED WITH THE PUMP AND COULD NOT BE EVALUATED. WE ARE ATTEMPTING TO CONTACT THE CUSTOMER/PT TO OBTAIN MORE INFO. IF MORE CUSTOMER/PT INFO BECOMES AVAILABLE, A FOLLOW UP REPORT MAY BE SUBMITTED.
REPORTED BY THE CUSTOMER AS: PUMP CONTINUED TO PUMP AFTER FORMULA WAS GONE. CUSTOMER STATED "PUMP IS FEEDING AIR TO PT." PT INJURY? NO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENTERALITE INFINITY ENTERAL FEEDING PUMP | LZH | MOOG MEDICAL DEVICES GROUP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |