FDA Adverse Event Malfunction Summary report: N

SUINSA (MARKETED BY CARESTREAM)

MDR report key: 1669061 · Received April 20, 2010

Report

Report Number
1000404222-2009-00002
Event Type
Malfunction
Date Received
April 20, 2010
Date of Event
August 11, 2009
Report Date
September 9, 2009
Manufacturer
SUINSA
Product Code
IYB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
RADIOLOGIC TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE STRAIN GAUGE (WHEATSTONE BRIDGE THAT MEASURES OPERATOR EFFORT) SIGNAL WILL BE EVALUATED PRIOR TO VERTICAL SERVO MOTOR ASSISTANCE IS CONNECTED/ENABLED. STRAIN GAUGE SIGNAL WILL BE NOW MEASURED BY A NEW REVISION OF THE BOARD THAT MONITORS THE ACTIVATION/DEACTIVATION OF THE OTC (OVERHEAD TUBE CRANE) HANDLE BUTTONS. AT THE MOMENT THAT ANY DEAD MAN BUTTON THAT DEMANDS VERTICAL ASSISTED MOVEMENT IS ENABLED BY OPERATOR ("ALL" OR "VERTICAL") IF THE STRAIN GAUGE EFFORT READOUT IS NOT 0 (OR CLOSE TO, DUE TO TOLERANCE BAND) SUCH BOARD WILL IGNORE SUCH BUTTON DEMAND. THEREFORE IF STRAIN GAUGE MALFUNCTION OCCURS (LIKE INCIDENT DESCRIBED), WHENEVER OPERATOR TRIES TO MOVE OTC VERTICAL AXIS BY DEPRESSING BUTTON, THE NEW REVISION BOARD WILL DETECT A NON CLOSE TO 0 EFFORT AND THEREFORE SUCH BUTTON DEPRESS WILL BE COMPLETELY IGNORED AND THE MOTOR SERVO ASSISTANCE WILL NOT BE CONNECTED AT ALL.

Description of Event or Problem · 1

THE TECHNICIAN WAS POSITIONING THE OTC (OVERHEAD TUBE CRANE) OVER A PT ON THE TABLE FOR A LATERAL EXAM. THE TECHNICIAN PRESSED THE "ALPHA" BUTTON, ALSO INADVERTENTLY PRESSING THE "ALL" BUTTON ON THE OTC OPERATOR HANDLE. THE OTC MOTOR DROVE THE OTC VERTICALLY DOWNWARD AT A HIGHER THAN USUAL RATE OF SPEED. THE OPERATOR DID NOT RELEASE THE "ALL" BUTTON QUICKLY ENOUGH TO PREVENT PT CONTACT. THE OTC CONTACTED THE PT WITH SOME AMOUNT OF FORCE IN THE HIP AREA. THE PT SAW A PHYSICIAN ON (B)(6) 2009. NO INJURY WAS DIAGNOSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUINSA (MARKETED BY CARESTREAM) TUBE MOUNT, X-RAY, DIAGNOSTIC, IYB IYB SUINSA NOVA

Patients

Seq Age Sex Outcome Treatment
1