SUINSA (MARKETED BY CARESTREAM)
Report
- Report Number
- 1000404222-2009-00002
- Event Type
- Malfunction
- Date Received
- April 20, 2010
- Date of Event
- August 11, 2009
- Report Date
- September 9, 2009
- Manufacturer
- SUINSA
- Product Code
- IYB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
THE STRAIN GAUGE (WHEATSTONE BRIDGE THAT MEASURES OPERATOR EFFORT) SIGNAL WILL BE EVALUATED PRIOR TO VERTICAL SERVO MOTOR ASSISTANCE IS CONNECTED/ENABLED. STRAIN GAUGE SIGNAL WILL BE NOW MEASURED BY A NEW REVISION OF THE BOARD THAT MONITORS THE ACTIVATION/DEACTIVATION OF THE OTC (OVERHEAD TUBE CRANE) HANDLE BUTTONS. AT THE MOMENT THAT ANY DEAD MAN BUTTON THAT DEMANDS VERTICAL ASSISTED MOVEMENT IS ENABLED BY OPERATOR ("ALL" OR "VERTICAL") IF THE STRAIN GAUGE EFFORT READOUT IS NOT 0 (OR CLOSE TO, DUE TO TOLERANCE BAND) SUCH BOARD WILL IGNORE SUCH BUTTON DEMAND. THEREFORE IF STRAIN GAUGE MALFUNCTION OCCURS (LIKE INCIDENT DESCRIBED), WHENEVER OPERATOR TRIES TO MOVE OTC VERTICAL AXIS BY DEPRESSING BUTTON, THE NEW REVISION BOARD WILL DETECT A NON CLOSE TO 0 EFFORT AND THEREFORE SUCH BUTTON DEPRESS WILL BE COMPLETELY IGNORED AND THE MOTOR SERVO ASSISTANCE WILL NOT BE CONNECTED AT ALL.
THE TECHNICIAN WAS POSITIONING THE OTC (OVERHEAD TUBE CRANE) OVER A PT ON THE TABLE FOR A LATERAL EXAM. THE TECHNICIAN PRESSED THE "ALPHA" BUTTON, ALSO INADVERTENTLY PRESSING THE "ALL" BUTTON ON THE OTC OPERATOR HANDLE. THE OTC MOTOR DROVE THE OTC VERTICALLY DOWNWARD AT A HIGHER THAN USUAL RATE OF SPEED. THE OPERATOR DID NOT RELEASE THE "ALL" BUTTON QUICKLY ENOUGH TO PREVENT PT CONTACT. THE OTC CONTACTED THE PT WITH SOME AMOUNT OF FORCE IN THE HIP AREA. THE PT SAW A PHYSICIAN ON (B)(6) 2009. NO INJURY WAS DIAGNOSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SUINSA (MARKETED BY CARESTREAM) | TUBE MOUNT, X-RAY, DIAGNOSTIC, IYB | IYB | SUINSA | NOVA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |