MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2023-182528
- Event Type
- Injury
- Date Received
- April 6, 2023
- Date of Event
- September 16, 2022
- Report Date
- November 23, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000414344
- PMA / PMN Number
- P160017
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
CUSTOMER RETURNED PUMP FOR AN ALLEGED NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM, EMS DISPATCHED/AMBULANCE/VISIT TO EMERGENCY ROOM AND WAS HOSPITALIZED FOR HIGH BGS/DKA FOUND ON 16-SEP-2022. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08675 INCHES. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. SUCCESSFULLY UPLOADED PUMP TO CARELINK. PLEASE SEE BELOW FOR THE DATE RANGE LISTED IN THE FORMATTED HISTORY FILE. THE FORMATTED HISTORY FILE LISTS DATA FROM 12/05/2022 TO 05/11/2023. THERE WAS NO DATA AVAILABLE TO VERIFY UNEXPECTED ALARMS/SUSPENDS AND BOLUS DELIVERY FOR THE EVENT DATE OF 16-SEP-2022. THERE WAS NO DATA AVAILABLE TO VERIFY UNEXPECTED PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE 16-SEP-2022 IN THE FORMATTED HISTORY FILE. THERE WAS NO DATA AVAILABLE TO VERIFY NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM IN THE FORMATTED HISTORY FILE ON THE EVENT DATE 16-SEP-2022. THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO UNEXPECTED OCCLUSIONS OR INSULIN FLOW BLOCKED ALARM NOTED DURING TESTING. THE PUMP WAS RECEIVED WITH THE ORIGINAL BATTERY CAP. NO CRACKED/DAMAGED BATTERY CAP NOTED DURING VISUAL INSPECTION. THE PUMP WAS RECEIVED WITHOUT A BATTERY INSTALLED. THE TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED BATTERY TUBE THREADS AND A SCRATCHED CASE. THE PUMP PASSED ALL THE REQUIRED TESTING. UNABLE TO VERIFY CUSTOMER ALLEGED FOR HIGH BGS/DKA. CUSTOMER ALLEGED FOR NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 500 MG/DL AND INSULIN FLOW WAS BLOCKED. THE CUSTOMER WAS TAKEN TO EMERGENCY ROOM AND ADMITTED TO THE HOSPITAL AND WAS TREATED WITH MANUAL INJECTION/INSULIN PEN. LENGTH OF HOSPITALIZATION OF 3 DAYS. TROUBLESHOOTING WAS PERFORMED AND IT WAS UNKNOWN WHETHER THE INSULIN PUMP WAS USED WITHIN 48 HOURS AND THE AUTO MODE FEATURE WAS ACTIVE AT THE TIME OF EVENT AND INSULIN FLOW BLOCK WAS RESOLVED BY COMPLETE SET CHANGE. IT WAS UNKNOWN WHETHER THE CUSTOMER WILL DISCONTINUE THE USE OF THE INSULIN PUMP. THE PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1852729 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | HG57795ZZ | 000000763000414344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Female | Other| H | FRN-MMT-332A-RSVR, UNOMED INF SET, OZP-MMT-7020MLA. |