POWERLINK SYSTEM
Report
- Report Number
- 2031527-2010-00043
- Event Type
- Injury
- Date Received
- April 27, 2010
- Date of Event
- March 26, 2010
- Report Date
- April 26, 2010
- Manufacturer
- ENDOLOGIX, INC.
- Product Code
- MIH
- PMA / PMN Number
- P040002
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
MODEL NO. 25-25-75L LOT NO. W10-0341-022 EXPIRATION DATE: 03/09/2013;MODEL NO. 25-25-95RL LOT NO. W09-1591-017 EXPIRATION DATE: 06/19/2012.REVIEW OF LOT RECORDS/WORK ORDERS, PRIOR REPORTS. NO ISSUES WERE NOTED. (B) (4) PATIENT HAD TWO 90 DEGREE ANGLES IN THE AORTIC NECK; ANATOMY IS OFF-LABEL PER INDICATIONS FOR USE.
PATIENT ANATOMY REPORTED WITH TWO 90° ANGLES IN THE AORTIC NECK (OFF-LABEL). THE PHYSICIAN HAD DIFFICULTY ADVANCING A TECH DEVICE STIFF WIRE INTO THE NECK DUE TO THE SEVERE ANGULATION. THE STIFF WIRE WAS TRADED OUT FOR A LUNDERQUIST. BALLOONING BILATERALLY WAS REQUIRED DUE TO NARROW, TORTUOUS ILIACS. THE 25-16-140BL BIFURCATED STENT GRAFT WAS DEPLOYED AND PLACED AT THE BIFURCATION. TWO 25-25-75L PROXIMAL EXTENSIONS WERE PLACED WITH GOOD OVERLAP OF THE BIFURCATED STENT GRAFT, BUT DURING PLACEMENT IN THE SEVERELY ANGULATED AORTA, COULD NOT GET SEAL BECAUSE THE FIRST EXTENSION SLIPPED WITHIN THE MAIN BODY OF THE BIFURCATED AND MOST PROXIMAL EXTENSION SLIPPED DOWN INTO THE DISTAL CUFF. THE PHYSICIAN WAS ABLE TO WEDGE THE MOST PROXIMAL EXTENSION INTO PLACE USING THE TIP OF THE DELIVERY SYSTEM. A 25-25-95RL SUPRARENAL PROXIMAL EXTENSION [W09-1591-017] WAS ADVANCED, BUT WHEN ATTEMPTING TO DEPLOY, IT SLIPPED DISTALLY ALSO DUE TO THE SEVERE ANGULATION. THIS EXTENSION WAS ALSO WEDGED INTO PLACE AS DESCRIBED PREVIOUSLY. A PALMAZ STENT WAS PLACED TO ATTEMPT TO SUPPORT THE SUPRARENAL EXTENSION AT THE RENALS. THE PHYSICIAN TRIED TO OBTAIN GUIDEWIRE ACCESS VIA A BRACHIAL APPROACH TO PLACE A THIRD EXTENSION, BUT WAS NOT ABLE TO ACCESS TO PLACE A WIRE. A 25-25-55L [W09-2904-015] WAS USED TO ATTEMPT TO BRIDGE THE GAP BETWEEN THE 25X75L AND 25X95RL EXTENSIONS. AFTER PLACING THE 4TH EXTENSION, THE SUPRARENAL EXTENSION SLIPPED DISTALLY BUT THE PALMAZ MAINTAINED ITS POSITION. AN ADDITIONAL 25-25-75L EXTENSION WAS ATTEMPTED TO BE PLACED JUST BELOW THE LEFT RENAL, COVERING THE RIGHT (LOWEST) RENAL BUT DUE TO THE SEVERE TORTUOSITY AND ANGULATION OF THE NECK WAS UNABLE TO GAIN APPOSITION. AT THAT TIME THE PHYSICIAN DECIDED TO CONVERT THE PATIENT TO OPEN REPAIR, REMOVING ALL EXTENSIONS AND LEAVING THE BIFURCATED STENT GRAFT IN PLACE. THE PATIENT WAS REPORTED TO HAVE BEEN DISCHARGED FROM THE HOSPITAL SINCE THE PROCEDURE AND IS DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERLINK SYSTEM | INFRARENAL PROXIMAL EXTENSION | MIH | ENDOLOGIX, INC. | 25-25-75L | W09-3308-008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |