BD BBL¿ SABOURAUD DEXTROSE AGAR WITH CHLORAMPHENICOL DEEP FILL
Report
- Report Number
- 1119779-2023-00388
- Event Type
- Malfunction
- Date Received
- April 4, 2023
- Date of Event
- March 13, 2023
- Report Date
- April 6, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JSI
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY DEAR VALUED CUSTOMER, THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT (B)(4) AGAINST SABOURAUD AGAR SABOURAUD AGAR W/CHMP, CATALOG NUMBER 254091, LOT NUMBER 3017543 WITH RESPECT TO PACKAGING ISSUE. EVENT DESCRIPTION: IT WAS REPORTED THAT AFTER OPENED THE BOXES (LOT 3017543), THE BAGS USED TO PRESERVE THE BOXES WERE OPEN AND THE AGARS WERE CONTAMINATED UPON RECEIPT. COMPLAINT HISTORY REVIEW: THE COMPLAINTS TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. SIMILAR COMPLAINTS WERE REPORTED DURING THAT PERIOD OF TIME ON THIS PRODUCT. HOWEVER, A TREND COULD NOT BE DETECTED. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. RETURNED SAMPLE ANALYSIS: RETAIN SAMPLES WERE REVIEWED, NEITHER CONTAMINATION NOR OPEN FOIL COULD BE IDENTIFIED. PHYSICAL SAMPLES WERE NOT RETURNED FOR EVALUATION. PICTURES WERE PROVIDED. BASED ON THE EVALUATION OF THE SHARED PICTURES CONTAMINATION AND PERFORATED FOIL COULD BE DETECTED. EVALUATION SUMMARY: AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL (STERILITY ASSURANCE LEVEL) CLAIM. IT IS FILLED ASEPTICALLY; THEREFORE, AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. ACCORDING TO OUR HIGH-QUALITY STANDARD, WE ONLY RELEASE PRODUCT BATCHES TO THE MARKET WITH AN AQL (ACCEPTABLE QUALITY LEVEL) = 0,65. ALTHOUGH THIS CONTAMINATION LEVEL IS VERY LOW, WE CANNOT COMPLETELY EXCLUDE THAT A CUSTOMER MAY RECEIVE A CONTAMINATED PLATE. UNCAREFUL HANDLING OF THE BOXES DURING THE TRANSPORT/STORAGE CAN PERFORATE THE FOIL. HOWEVER, A DEFINITE ROOT CAUSE COULD NOT BE DETERMINED. INVESTIGATION CONCLUSION: BASED ON THE ABOVE-MENTIONED EVALUATION THE COMPLAINT CAN BE CONFIRMED. A DEFINITE ROOT CAUSE COULD NOT BE DETERMINED. BD REGRETS THE INCONVENIENCES YOU HAVE EXPERIENCED AND WILL CONTINUE TO MONITOR SIMILAR INCOMING COMPLAINTS.
D.3 COMMON DEVICE NAME: CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL. E.6 INITIAL REPORTER E-MAIL: (B)(6). G5. THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US, BUT IT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ SABOURAUD DEXTROSE AGAR WITH CHLORAMPHENICOL DEEP FILL, CATALOG NUMBER 221851 WITH 510K #EXEMPT. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING THE BD BBL¿ SABOURAUD DEXTROSE AGAR WITH CHLORAMPHENICOL DEEP FILL THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AGARS WERE CONTAMINATED UPON RECEIPT. THE CUSTOMER RECEIVED THE AGARS AND REALIZED THAT THEY WERE CONTAMINATED ON (B)(6) 2023 BEFORE USE.
IT WAS REPORTED THAT WHILE USING THE BD BBL¿ SABOURAUD DEXTROSE AGAR WITH CHLORAMPHENICOL DEEP FILL THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AGARS WERE CONTAMINATED UPON RECEIPT THE CUSTOMER RECEIVED THE AGARS AND REALIZED THAT THEY WERE CONTAMINATED ON 03/13/23 BEFORE USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387071 | BD BBL¿ SABOURAUD DEXTROSE AGAR WITH CHLORAMPHENICOL DEEP FILL | SEE H.10 | JSI | BECTON, DICKINSON & CO. (SPARKS) | 3017543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |