FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 16675064
·
Received April 4, 2023
Report
- Report Number
- 3013756811-2023-48239
- Event Type
- Malfunction
- Date Received
- April 4, 2023
- Date of Event
- March 15, 2023
- Report Date
- April 4, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613793
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. SYSTEM CHECK WAS PERFORMED WITH TANDEM TECHNICAL SUPPORT (TTS) AND NO ISSUES WERE IDENTIFIED. CUSTOMER CHANGED THE PUMP SUPPLIES AND SUCCESSFULLY RESUMED INSULIN THERAPY. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS 200-430 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1130375 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000354 | 00850006613793 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Male | INFUSION SET: AUTOSOFT 90INSULIN TYPE: HUMALOG |