UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2010-00220
- Event Type
- Malfunction
- Date Received
- April 26, 2010
- Date of Event
- April 5, 2010
- Report Date
- April 26, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE SYSTEM IS ROUTINELY CALIBRATED EVERY 24 HOURS AND QC SAMPLES ARE RUN EVERY 8 HOURS. PRIOR TO THE EVENT, QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) REPLACED THE FLOW CELL. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY HIGH SODIUM (NA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS. UPON REPEAT LOWER RESULTS WERE OBTAINED. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LAB. THE CUSTOMER HAS NOT RECEIVED ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED TO OR CONNECTED WITH THIS EVENT.