FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS

MDR report key: 1667224 · Received April 26, 2010

Report

Report Number
2050012-2010-00231
Event Type
Malfunction
Date Received
April 26, 2010
Date of Event
April 1, 2010
Report Date
April 26, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM IS CALIBRATED EVERY 24 HOURS AND QC SAMPLES ARE RUN EVERY 8 HOURS. PRIOR TO THE EVENT, QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) REPLACED SEVERAL HARDWARE COMPONENTS. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY HIGH SODIUM (NA) AND CALCIUM (CA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS. UPON REPEAT LOWER RESULTS WERE OBTAINED. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE OF THE LAB. THERE WAS NO AFFECT TO PATIENT OR USER ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS CLINICAL CHEMISTRY ANALYZER JGS BECKMAN COULTER INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1