UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS
MDR report key: 1667224
·
Received April 26, 2010
Report
- Report Number
- 2050012-2010-00231
- Event Type
- Malfunction
- Date Received
- April 26, 2010
- Date of Event
- April 1, 2010
- Report Date
- April 26, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE SYSTEM IS CALIBRATED EVERY 24 HOURS AND QC SAMPLES ARE RUN EVERY 8 HOURS. PRIOR TO THE EVENT, QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) REPLACED SEVERAL HARDWARE COMPONENTS. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY HIGH SODIUM (NA) AND CALCIUM (CA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS. UPON REPEAT LOWER RESULTS WERE OBTAINED. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE OF THE LAB. THERE WAS NO AFFECT TO PATIENT OR USER ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JGS | BECKMAN COULTER INC. | DXC 800 PRO | NA |