GRAFTS ADVANTA VXT W/GDS
Report
- Report Number
- 3011175548-2023-00090
- Event Type
- Malfunction
- Date Received
- April 3, 2023
- Report Date
- May 26, 2023
- Manufacturer
- ATRIUM MEDICAL CORPORATION
- Product Code
- DSY
- UDI-DI
- 00650862221152
- PMA / PMN Number
- K992960
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- 003
Narratives
UPON COMPLETION OF THE INVESTIGATION INTO THIS EVENT A FOLLOW UP REPORT WILL BE SUBMITTED.
RELATED MDR: 3011175548-2023-00091 . ADDITIONAL INFORMATION: SECTION D9 & H6. INVESTIGATION: THE USER REPORTED THAT DURING A REVIEW OF INVENTORY THEY DISCOVERED TWO DEVICES WHERE THE THICKNESS OF THE GRAFT WAS NOT CONSISTENT. PHOTOS WERE PROVIDED AND SHOW THE CONCERN IS WITH VARIATION IN WALL THICKNESS AT THE END OF THE GRAFTS (ONE SIDE APPEARS THICKER THAN THE OTHER). THE PRODUCT WAS IN INVENTORY AT THE TIME OF THE OBSERVATION AND WAS NOT USED IN ANY PROCEDURE OR ON A PATIENT. THE INVENTORY REVIEW WAS CONDUCTED BY THE USER DUE TO THE OBSERVATION OF A GRAFT WITH VARIABLE WALL THICKNESS DURING A PROCEDURE (REFERENCE COMPLAINT (B)(4)). THE TWO DEVICES FOUND IN THE INVENTORY REVIEW WERE RETURNED FOR EVALUATION. THE RETURNED DEVICES WERE EVALUATED AND ARE FINISHED GOOD CODE 22115 - 4-7X45 SW SH ADVANTA VXT GDS WITH PRODUCT LOT NUMBER 486233 AND 487533. THE ENTIRE UNUSED DEVICES WERE RETURNED FOR EVALUATION AND SHOWED NO SIGNS OF HAVING EVER ENTERED THE SURGICAL FIELD. THE RETURNED GRAFTS HAD NO VISIBLE DAMAGE NOTED DURING THE VISUAL INSPECTION OF THE DEVICE AND THE WALL THICKNESS WAS NOT CONCENTRIC AROUND THE CIRCUMFERENCE OF THE GRAFT. THE TACTILE INSPECTION DID IDENTIFY THAT THE GRAFT DID NOT HAVE A CONSISTENT FEELING ALONG THE LENGTH AND CIRCUMFERENCE DUE TO THE WALL THICKNESS NOT BEING CONCENTRIC. THE DESCRIPTION OF THE EVENT SPECIFIED THAT THERE APPEARED TO BE A WALL THICKNESS DISCREPANCY OF THE GRAFT FROM ONE SIDE TO THE OTHER. THIS HAS BEEN CONFIRMED AS THIS CAN BE SEEN VISUALLY AND FELT TACTICALLY. REPEATED PUNCTURE CHARACTERISTIC TESTING AND SUTURE RETENTION TESTING WERE CONDUCTED ON THE RETURNED PRODUCT. ALL RESULTS OBTAINED PASSED THE REQUIREMENTS. THE WALL THICKNESS OF THE GRAFT WAS ALSO MEASURED. MEASUREMENTS WERE TAKEN IN 4 LOCATIONS AROUND THE CIRCUMFERENCE INCLUDING THE THINNEST VISIBLE WALL OF THE GRAFT ON BOTH GRAFTS SAMPLES. THE TOP ASSEMBLY PART SPECIFICATION 001095-XXX-XXX TOP, VXT, SW, STPR REVISION AN CALLS OUT THAT THE SMALL END OF THE GRAFT HAVE A WALL THICKNESS RANGING FROM .020 TO .029 INCHES AND THE LARGE END HAVE A WALL THICKNESS RANGE WITHIN .016 TO .028 INCHES. THE THINNEST WALL THICKNESS OBSERVED WAS ON THE LARGE END OF THE GRAFT AND WAS RECORDED AS .0111 INCHES. THE NEXT THINNEST MEASUREMENT WAS .0123 INCHES ALSO TAKEN ON THE LARGE END OF THE GRAFT. IN TOTAL THERE WERE FIVE WALL THICKNESS MEASUREMENTS OF THE EIGHT RECORDED THAT WERE BELOW THE .016 INCH MINIMUM REQUIREMENT OF THE LARGE END OF THE GRAFT. THERE WAS ONE MEASUREMENT BELOW THE .020 INCH MINIMUM REQUIREMENT ON THE SMALL END OF THE GRAFT AND WAS .0193 INCH. THE WALL THAT WAS VISUALIZED AS THINNER WAS CONFIRMED TO BE THINNER WITH THE WALL THICKNESS MEASUREMENTS, FURTHER CONFIRMING THE COMPLAINT. THE WALL THICKNESS OF THE GRAFT IN QUESTION IS NOT CONFORMING TO THE PART SPECIFICATION DRAWING 001095-XXX-XXX AND THE GRAFTS ARE CONSIDERED NONCONFORMING TO THE WALL THICKNESS REQUIREMENTS . A REVIEW OF THE FINISHED GOOD LOT DEVICE HISTORY RECORDS AS WELL AS ALL SUB ASSEMBLY LOTS SHOW THAT THE GRAFTS PASSED ALL QUALITY AND PERFORMANCE REQUIREMENTS DURING THE MANUFACTURE OF THE PRODUCT. THERE WERE NO NONCONFORMITIES IDENTIFIED WITHIN THE DHR THAT COULD BE RELATED TO THE COMPLAINT AND THERE IS NO EVIDENCE THAT PROCEDURAL OR SPECIFICATION CHANGES THAT OCCURRED SURROUNDING THE TIME OF THE BUILD COULD BE RELATED TO THE COMPLAINT. IT IS IMPORTANT TO NOTE THAT WALL THICKNESS IS NOT MEASURED AT THE TOP ASSEMBLY OR FINISHED GOOD LEVEL. A COMPLAINT HISTORY, COMPLAINT TREND, RECURRING LOT NUMBER AND CAPA REVIEW DID IDENTIFY FIVE RELATED COMPLAINTS FOR THE SAME PRODUCT CODE. FOUR OF THE FIVE COMPLAINTS WERE RECEIVED FROM THE SAME INSTITUTION THAT REPORTED THIS COMPLAINT. ALL THE COMPLAINTS ARE CONFIRMED AND THE ROOT CAUSE WAS LISTED AS USER PREFERENCE, MANUFACTURING OR HAS NOT BEEN DETERMINED YET (THE INVESTIGATION IS ONGOING). THE INVESTIGATION FOR THIS COMPLAINT IDENTIFIES THAT THE MANUFACTURING PROCESS ALLOWS VARIATION IN THE WALL THICKNESS. HOWEVER, THE WALL THICKNESS MEASUREMENTS OBTAINED ON THE PARTS DO NOT MEET THE PART SPECIFICATION 001095-XXX-XXX TOP, VXT, SW, STPR REVISION AN FOR WALL THICKNESS. THERE IS NO IN PROCESS CONTROL TO DETECT GRAFTS WITH NON CONFORMING WALL THICKNESSES. THE TACTILE INSPECTION PERFORMED ON THE RETURNED DEVICES ALSO INDICATED THAT THEY SHOULD NOT HAVE PASSED THE TACTILE INSPECTION PERFORMED IN PROCESS PER (B)(4). ADDITIONALLY, THERE WAS INFORMATION OBTAINED DURING THE INVESTIGATION THAT THE SPECIFICATIONS ON THE DRAWING MAY NOT BE COMPARATIVE WITH THE WALL THICKNESS RESULTS OBTAINED FROM THE DROP GAUGE WALL THICKNESS TEST METHOD MEASUREMENTS VALIDATED FOR TOP ASSEMBLY THICKNESS, POINTING TO A POTENTIAL DEFICIENCY IN THE DESIGN REQUIREMENTS. IN THIS REGARD THE ROOT CAUSE IS RELATED TO MANUFACTURING WITH A CONTRIBUTING FACTOR RELATED TO DESIGN. A RISK REVIEW FOUND THAT THE RISK MANAGEMENT DOCUMENTS FOR THIS PRODUCT ADEQUATELY ADDRESS THE REPORTED DEFECT AND THE SEVERITY AND ANTICIPATED OCCURRENCE LEVEL ARE APPROPRIATE. THE LABELING REVIEW DETERMINED THAT THE INSTRUCTIONS FOR USE ARE APPROPRIATE AS THERE ARE CLEAR WARNINGS AND USABILITY INSTRUCTIONS RELATED TO USE OF A DEVICE THAT APPEARS DAMAGED. AS THE WALL THICKNESS OF THE GRAFT IN QUESTION IS NOT CONFORMING TO THE PART SPECIFICATION DRAWING 001095-XXX-XXX DOES NOT PASS THE TACTILE INSPECTION, AND THERE IS INFORMATION SUGGESTING THAT THE SPECIFICATIONS ON THE DRAWING MAY NOT BE COMPARATIVE WITH THE WALL THICKNESS RESULTS OBTAINED FROM THE DROP GAUGE WALL THICKNESS TEST METHOD MEASUREMENTS, THIS COMPLAINT AND THE ASSOCIATED COMPLAINT WILL BE ESCALATED TO A CORRECTIVE ACTION REQUEST (CR 803795).
AFTER OPENING, THE THICKNESS OF THE GRAFT WAS NOT CONSISTENT, WE REQUESTED A RETURN.
N/A
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1265779 | GRAFTS ADVANTA VXT W/GDS | PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER | DSY | ATRIUM MEDICAL CORPORATION | 22115 | 486233 | 00650862221152 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |