FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T (EUA)

MDR report key: 16663169 · Received April 3, 2023

Report

Report Number
1221359-2023-00570
Event Type
Malfunction
Date Received
April 3, 2023
Date of Event
February 26, 2023
Report Date
June 1, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
00811877011354
PMA / PMN Number
EUA210517
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED; SINGLE-USE DEVICE.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1093518 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 192-000 / LOT 1093518 AND TEST BASE PART NUMBER 192-430 / LOT 1093518. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED, UNCONFIRMED AND CONFLICTING RESULTS) RELATED TO KIT LOT 1093518 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE AND CROSS-CONTAMINATION. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. H3 OTHER TEXT : DEVICE DISCARDED; SINGLE-USE DEVICE.

Description of Event or Problem · 0

THE CONSUMER REPORTED CONFLICTING RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023. PER THE CONSUMER, THE FIRST TEST TAKEN GENERATED A POSITIVE RESULT AND THE SECOND TEST CONDUCTED THE SAME DAY AT DIFFERENT (B)(6) WHICH GENERATED NEGATIVE RESULT. CONFIRMATION TESTING WAS NOT PERFORMED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED CONFLICTING RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023. PER THE CONSUMER, THE FIRST TEST TAKEN GENERATED A POSITIVE RESULT AND THE SECOND TEST CONDUCTED THE SAME DAY AT DIFFERENT WALGREEN WHICH GENERATED NEGATIVE RESULT. CONFIRMATION TESTING WAS NOT PERFORMED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581327 ID NOW COVID-19 2.0 TEST KIT 24T (EUA) REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1093518 00811877011354

Patients

Seq Age Sex Outcome Treatment
1 Unknown