ID NOW COVID-19 2.0 TEST KIT 24T (EUA)
Report
- Report Number
- 1221359-2023-00562
- Event Type
- Malfunction
- Date Received
- April 3, 2023
- Date of Event
- February 26, 2023
- Report Date
- June 1, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 00811877011354
- PMA / PMN Number
- EUA210517
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
TESTING WAS PERFORMED IN DUPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1093518 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 192-000/LOT 1093518 AND TEST BASE PART NUMBER 192-430/LOT 1093518. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED, UNCONFIRMED AND CONFLICTING RESULTS) RELATED TO KIT LOT 1093518 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE AND CROSS-CONTAMINATION. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. H3 OTHER TEXT: DEVICE DISCARDED; SINGLE-USE DEVICE.
THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED; SINGLE-USE DEVICE.
THE CUSTOMER REPORTED NINE (9) UNCONFIRMED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED BETWEEN (B)(6) 2023 AND (B)(6) 2023. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF NINE (9). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 USING KITTED SWAB ON A NASAL SAMPLE. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
THE CUSTOMER REPORTED NINE (9) UNCONFIRMED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED BETWEEN 02FEB2023 AND 27MAR2023. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF NINE (9). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 USING KITTED SWAB ON A NASAL SAMPLE. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1265477 | ID NOW COVID-19 2.0 TEST KIT 24T (EUA) | REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1093518 | 00811877011354 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |