FDA Adverse Event Malfunction Summary report: N

COBAS INTEGRA 400 PLUS

MDR report key: 1666226 · Received April 23, 2010

Report

Report Number
1823260-2010-02449
Event Type
Malfunction
Date Received
April 23, 2010
Date of Event
April 8, 2010
Report Date
April 23, 2010
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K951595
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

THE USER RECEIVED ERRONEOUS RESULTS FOR FIVE PATIENT SAMPLES. OF THE DATA PROVIDED, THE RESULTS FOR THREE SAMPLES WERE DISCREPANT. ALL REPEAT TESTING WAS PERFORMED ON (B) (6) 2010. SAMPLE 1 INITIAL SODIUM RESULT WAS 134 MMOL/L, REPEATED RESULTS WERE 129, 114, 135, 135, 135, 134 AND 128 MMOL/L. SAMPLE 2 INITIAL CALCIUM RESULT WAS 10.14 MG/DL, REPEAT RESULTS WERE 9.25, 9.26, 9.20, 9.35 AND 9.27 MG/DL. INITIAL GLUCOSE RESULT WAS 298 MG/DL, REPEAT RESULTS WERE 132, 132, 133, 131, 132, 132, 132, 133, 133 AND 132 MG/DL. SAMPLE 3 INITIAL CALCIUM RESULT WAS 10.12 MG/DL, REPEAT RESULTS WERE 9.22, 9.27, 9.22, 9.28, 9.30 MG/DL. INITIAL GLUCOSE RESULT WAS 196 MG/DL, REPEAT RESULTS WERE 88, 87, 87, 87, 88, 88, 88, 87, 87 AND 87 MG/DL. THE ERRONEOUS RESULTS WERE REPORTED. THE USER DID NOT KNOW IF ANY PATIENTS WERE TREATED AND HAD NO ACCESS TO INFORMATION REGARDING PATIENT CURRENT CONDITION. THE LOT NUMBER OF THE SODIUM ELECTRODE WAS NOT PROVIDED. THE REAGENT LOT NUMBERS WERE CALCIUM 61691701 AND GLUCOSE 61774601. THE FIELD SERVICE REPRESENTATIVE DETERMINED THERE WAS A PROBLEM WITH DOSAGE B PIPETTE. HE PERFORMED A CHECK TEST, REPLACED THE DOSAGE PIPETTE B AND SWITCHED THE DOSAGE VALVES. TO VERIFY THE ANALYZER OPERATION, HE RAN A 10 POINT PRECISION TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS INTEGRA 400 PLUS CLINICAL CHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1