630G INSULIN PUMP MMT-1715K 630G BLACK MG
Report
- Report Number
- 2032227-2023-180105
- Event Type
- Injury
- Date Received
- April 3, 2023
- Date of Event
- March 10, 2023
- Report Date
- May 10, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 000000763000316631
- PMA / PMN Number
- P150001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- 003
Narratives
MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. INFORMATION HAS BEEN CORRECTED WHICH WAS NOT CORRECT IN THE INITIAL REPORT. THE INFORMATION HAS BEEN PROVIDED IN SECTION B5 WITH THIS REPORT.
(B)(4). THE TEST P-CAP LOCKS PROPERLY IN PLACE IN THE RESERVOIR COMPARTMENT NOTED. PUMP PILLOWING KEYPAD OVERLAY AND CRACKED SELECT BUTTON KEYPAD OVERLAY DURING THE VISUAL INSPECTION. THUS AND CARELINK SOFTWARE WAS UTILIZED AND DOWNLOADED TRACE/HISTORY FILES PROPERLY. IN FURTHER FULL REVIEW OF THE PUMP HISTORY ON THE EVENT DATE OF (B)(6) 2023, FOUND NO BOLUS DELIVERY AND THE DAILY TOTAL OF BOLUS INSULIN DELIVERED ARE 0 U. PUMP PASSED THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.0871 INCHES. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO MOISTURE OR COMPONENT DAMAGE ON THE ELECTRONICS, FORCE SENSOR AND MOTOR ASSEMBLY NOTED. THE FORCE SENSOR OFFSET MEASURED WITHIN SPEC RANGE DURING TESTING (23.7 MV). IN SUMMARY, PUMP PASSED ALL REQUIRED TESTING. UNABLE TO VERIFY CUSTOMER COMPLAINT FOR LOW BG. CUSTOMER ALLEGED FOR THE PUMP OVER DELIVERY WAS NOT CONFIRMED. THE FORCE SENSOR IS WITHIN SPECIFICATION. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
UPDATE: CUSTOMER REPORTED THAT SHE WAS HAVING LOW BLOOD SUGAR WHEN SHE BOLUSES. CUSTOMER IS ON TRULICITY. SHE HAS BEEN EXPERIENCING LOW FOR 4-5 MONTHS SINCE SHE STARTED TRULICITY. CUSTOMER ALSO SAID INSULIN TYPE/CONCENTRATION WAS NOT CORRECT. PARAMEDICS WERE CALLED ON (B)(6) 2023 FOR LOW BLOOD GLUCOSE OF 43MG/DL. CUSTOMER TREATED WITH FOOD. CUSTOMER ALLEGED OVER DELIVERY. PUMP WAS USED AT THE TIME OF THE INCIDENT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED HYPOGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 43 MG/DL. THE CUSTOMER WAS ATTENDED BY THE EMERGENCY MEDICAL SERVICES AND TREATED FOR HYPOGLYCEMIA WITH FOOD. TROUBLESHOOTING WAS PERFORMED AND THE CUSTOMER ALLEGES THAT THE INSULIN PUMP WAS OVERDELIVERING THE INSULIN. THE CUSTOMER WAS USING THE INSULIN PUMP SYSTEM WITHIN 48 HOURS OF THE REPORTED LOW BLOOD GLUCOSE EVENT. THE CUSTOMER WILL DISCONTINUE USING THE DEVICE AND THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1266299 | 630G INSULIN PUMP MMT-1715K 630G BLACK MG | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715K | HG4906H | 000000763000316631 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Female | Required Intervention | FRN-MMT-332A - RSVR, UNOMED SET |