FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 1666133 · Received April 23, 2010

Report

Report Number
2954323-2010-00548
Event Type
Malfunction
Date Received
April 23, 2010
Date of Event
March 27, 2010
Report Date
April 23, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S METER (B)(4) WAS RETURNED AND TESTED WITH RETAINED TEST STRIPS LOT # 1002834.THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED READINGS OF 376 MG/DL AND 101 MG/DL WERE FOUND IN THE DEVICE'S INTERNAL MEMORY LOG WITHIN 10 MINUTES. NOTE: ADDITIONAL EVALUATION RESULTS: (DEVICE PERFORMED ACCORDING TO SPECIFICATIONS)

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 376 MG/DL AND 101 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.