FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-00548
- Event Type
- Malfunction
- Date Received
- April 23, 2010
- Date of Event
- March 27, 2010
- Report Date
- April 23, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER'S METER (B)(4) WAS RETURNED AND TESTED WITH RETAINED TEST STRIPS LOT # 1002834.THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED READINGS OF 376 MG/DL AND 101 MG/DL WERE FOUND IN THE DEVICE'S INTERNAL MEMORY LOG WITHIN 10 MINUTES. NOTE: ADDITIONAL EVALUATION RESULTS: (DEVICE PERFORMED ACCORDING TO SPECIFICATIONS)
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 376 MG/DL AND 101 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.