FDA Adverse Event
Malfunction
Summary report: N
ADVANCE 3000 W/SCALE, ADVANCE FRAME
MDR report key: 1665999
·
Received April 16, 2010
Report
- Report Number
- 1824206-2010-04095
- Event Type
- Malfunction
- Date Received
- April 16, 2010
- Date of Event
- January 7, 2008
- Report Date
- January 7, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ACCOUNT FOUND LOOSE CONNECTIONS INSIDE THE FOOT PANEL, ACCOUNT TIGHTENED THE CONNECTIONS AND ISSUE IS RESOLVED. PURSUANT TO OUR RESPONSE TO WARNING LETTER 2008-DT-04, HILL-ROM IS CONTINUING ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY. THIS EFFORT WILL CONTINUE UNTIL WE ARE CURRENT.
Description of Event or Problem · 1
ACCOUNT ALLEGED THE BED HAS A LOSS OF GROUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANCE 3000 W/SCALE, ADVANCE FRAME | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM RITTER | P1165F55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |