CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2023-12012
- Event Type
- Injury
- Date Received
- March 31, 2023
- Date of Event
- February 27, 2023
- Report Date
- May 24, 2023
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- UDI-DI
- 00690103186260
- PMA / PMN Number
- P860057
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
STENOSIS, WHICH DEVELOPS PROGRESSIVELY OVER TIME, CAN BE DUE TO A NUMBER OF ISSUES. ADDITIONALLY, THERE CAN BE A NUMBER OF POTENTIAL KNOWN AND UNKNOWN PATIENT-RELATED CONTRIBUTING FACTORS. STRUCTURAL VALVE DETERIORATION (SVD) IS THE MOST COMMON REASON FOR BIOPROSTHESIS EXPLANT AND ENCOMPASSES MULTIPLE FAILURE MODES, INCLUDING CALCIFICATION, NON-CALCIFIC DEGENERATION, DEHISCENCE, CUSP THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. SUCH FAILURE MODES, OCCURRING SINGULARLY OR CONCOMITANTLY, MAY CONTRIBUTE TO STENOSIS AND/OR REGURGITATION. ALTERNATIVELY, NON-STRUCTURAL VALVE DYSFUNCTION (NSVD) MAY ALSO PLAY A ROLE IN THE DEVELOPMENT OF VALVULAR STENOSIS. STENOSIS IS MOST COMMONLY RELATED TO PATIENT FACTORS AND IS NOT USUALLY AN INDICATION OF A DEVICE MALFUNCTION. THE ROOT CAUSE OF THIS EVENT WAS DETERMINED TO BE DUE TO PATIENT RELATED FACTORS. THE STENOSIS IN THIS CASE WAS IMPACTED BY THE PROGRESSION OF THE PATIENT'S UNDERLYING VALVULAR DISEASE PATHOLOGY WITH OR WITHOUT STRUCTURAL VALVE DETERIORATION AND/OR NONSTRUCTURAL DYSFUNCTION. THE SUBJECT DEVICE IS NOT AVAILABLE FOR EVALUATION AS IT REMAINS IMPLANTED IN THE PATIENT. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED. H11: CORRECTED DATA D8 - "NO" FIELD SHOULD HAVE BEEN SELECTED ON INITIAL FDA REPORT. E1 (STATES LIST) - GEORGIA SHOULD HAVE BEEN SELECTED ON INITIAL FDA REPORT.
THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS PRIOR TO RELEASE FOR DISTRIBUTION. THERE WERE NO ISSUES IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, A CONCLUSION HAS YET TO BE ESTABLISHED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY UPON INVESTIGATION COMPLETION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.
THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, A CONCLUSION HAS YET TO BE ESTABLISHED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY UPON INVESTIGATION COMPLETION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 27MM 7300TFX MITRAL VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 5 YEARS, 8 MONTHS DUE TO STENOSIS. THE PATIENT PRESENTED WITH HEART FAILURE. THE PROCEDURE WAS PERFORMED WITH A 29MM 9600TFX TRANSCATHETER VALVE. THE PATIENT WAS STABLE POST PROCEDURE.
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 27MM 7300TFX MITRAL VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 5 YEARS, 8 MONTHS DUE TO UNKNOWN REASON. THE PROCEDURE WAS PERFORMED WITH A 29MM 9600TFX TRANSCATHETER VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1720321 | CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART VALVE | DYE | EDWARDS LIFESCIENCES | 7300TFX | 00690103186260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Male | Required Intervention| L| H |